DOI 10.1016/j.jse.2018.11.055

Медицинская библиотека · Journal of shoulder and elbow surgery · 2019

Biomechanical comparison of 2 augmented glenoid designs: an integrated kinematic finite element analysis.

Sabesan Vani J, Lima Diego J L, Whaley James D, Pathak Varun, Zhang Liying

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Augmented glenoid implants are available to help restore the biomechanics of the glenohumeral joint with excessive retroversion. It is imperative to understand their behavior to make a knowledgeable preoperative decision. Therefore, our goal was to identify an optimal augmented glenoid design based on finite element analysis (FEA) under maximum physiological loading.

METHODS

FEA models of 2 augmented glenoid designs-wedge and step-were created per the manufacturers' specifications and virtually implanted in a scapula model to correct 20° of retroversion. Simulation of shoulder abduction was performed using the FEA shoulder model. The glenohumeral force ratio, relative micromotion, and stress levels on the cement mantle, glenoid vault, and backside of the implants were compared between the 2 designs.

RESULTS

The force ratio was 0.56 for the wedge design and 0.87 for the step design. Micromotion (combination of distraction, translation, and compression) was greater for the step design than the wedge design. Distraction measured 0.05 mm for the wedge design and 0.14 mm for the step component. Both implants showed a similar pattern for translation; however, compression was almost 3 times greater for the step component. Both implants showed high stress levels on the cement mantle. At the glenoid vault and on the implants, the stress levels were 1.65 MPa and 6.62 MPa, respectively, for the wedge design and 3.78 MPa and 13.25 MPa, respectively, for the step design.

CONCLUSION

Implant design slightly affects joint stability; however, it plays a major role regarding long-term survival. Overall, the augmented wedge design provides better implant fixation and stress profiles with less micromotion.

Для цитирования: Sabesan Vani J, Lima Diego J L, Whaley James D, Pathak Varun, Zhang Liying. Biomechanical comparison of 2 augmented glenoid designs: an integrated kinematic finite element analysis.. Journal of shoulder and elbow surgery. 2019. DOI: 10.1016/j.jse.2018.11.055.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.