Development of a latex microsphere-based lateral flow immunoassay for the diagnosis of schistosomiasis japonica.

Tabilin Emmanuel John, Gordon Catherine A, Mu Yi, Jiz Mario, Inobaya Marianette, Avenido-Cervantes Eleonor, Gray Darren, Mationg Mary Lorraine, McManus Donald P, Egwang Thomas G, Adriko Moses, Sako Yasuhito, Sato Marcello Otake, Sato Megumi, You Hong, Kelly Matthew, Cai Pengfei

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Zoonotic schistosomiasis, caused by Schistosoma japonicum, is prevalent in China, the Philippines and Indonesia. Rapid point-of-care (POC) diagnostics are attractive and promising tools for evaluating the efficacy of intervention strategies for schistosomiasis control.

METHODOLOGY: The diagnostic potential of five recombinant antigens was tested by enzyme-linked immunosorbent assay (ELISA) using sera from individuals with positive Kato-Katz (KK) results for S. japonicum (n = 28) and non-endemic controls (n = 12). A latex microsphere (LM)-based lateral flow immunoassay (LFIA) incorporating the recombinant SjSAP4 (rSjSAP4) was developed for the diagnosis of schistosomiasis japonica. The test conditions including diluent, dilution factor and reaction time, were optimised for the developed LFIA. Under the optimised conditions, serum samples from individuals living in a barangay endemic for S. japonicum (n = 549) and non-endemic controls (n = 50) were tested with the established LFIA cassettes. The results were imaged by a smartphone and analysed by the ImageJ program. The intensity ratio of the test line to the control line (T/C ratio) was calculated for each cassette.

MAIN FINDINGS: ELISA confirmed that rSjSAP4 was the optimal candidate for serological diagnosis of schistosomiasis japonica. Under optimal testing conditions, the developed LFIA strips had a sensitivity of 80.6% and a specificity of 98.0% at a cut-off T/C ratio of 0.1031. Moreover, the results of the LM-based LFIA was positively correlated with those obtained from the rSjSAP4-ELISA (r = 0.8270, 95% CI, 0.7990-0.8514; p < 0.0001). The schistosomiasis prevalence determined by the LFIA strips was about 1.8 times greater than that obtained with the 6-slide KK procedure performed on three stool samples.

CONCLUSIONS/SIGNIFICANCE: The developed LFIA represents a POC diagnostic tool that is suitable for onsite screening of human S. japonicum infection with minimal equipment needed. The established immunochromatographic assay complies with most of the WHO's ASSURED criteria for POC diagnostics.

Для цитирования: Tabilin Emmanuel John, Gordon Catherine A, Mu Yi, Jiz Mario, Inobaya Marianette, Avenido-Cervantes Eleonor, Gray Darren, Mationg Mary Lorraine, McManus Donald P, Egwang Thomas G, Adriko Moses, Sako Yasuhito, Sato Marcello Otake, Sato Megumi, You Hong, Kelly Matthew, Cai Pengfei. Development of a latex microsphere-based lateral flow immunoassay for the diagnosis of schistosomiasis japonica.. PLoS neglected tropical diseases. 2024. DOI: 10.1371/journal.pntd.0012742.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.