DOI 10.1016/j.jcin.2018.05.025

Медицинская библиотека · JACC. Cardiovascular interventions · 2018

WIRION Embolic Protection System in Lower Extremity Arterial Interventions: Results of the Pivotal WISE LE Trial.

Shammas Nicolas W, Pucillo Anthony, Jenkins J Stephen, Garcia Lawrence A, Davis Tom, Aronow Herbert D, Armstrong Ehrin J, Zeller Thomas, Scheinert Dierk, Rosenschein Uri, Gray William

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

The WISE LE (WIRION™ EPS in Lower Extremities Arteries) study was designed to assess the clinical performance of the WIRION Embolic Protection System (EPS) in subjects undergoing lower extremity atherectomy for the treatment of peripheral artery disease.

BACKGROUND

Embolization is ubiquitous during endovascular procedures for lower extremity peripheral artery disease.

METHODS

The WISE LE was a multicenter study, performed in the United States and Germany. The primary endpoint was freedom from major adverse events (MAEs) occurring within 30 days post-procedure and was compared with an objective performance goal derived from historical atherectomy trials. MAE was defined as a serious adverse event that resulted in death, acute myocardial infarction, thrombosis, pseudoaneurysm, dissection (grade C or greater), or clinical perforation at the filter location, clinically relevant distal embolism, unplanned amputation, or clinically driven target vessel revascularization. The study also included a histopathological analysis of debris captured by the filter during the procedures.

RESULTS

The study protocol specified enrollment of 153 patients with the primary endpoint successfully met if 18 (12.0%) or fewer MAEs occurred. A pre-specified interim analysis performed after 103 patients revealed only 2 MAEs, and the study was stopped because it had met its pre-determined metric for success. Lesion deemed not accessible by the WIRION EPS occurred in 7 patients. Debris of <1-mm, 1- to 2-mm, and >2-mm diameter were found in 98%, 22%, and 9% of patients, respectively.

CONCLUSIONS

The WIRION EPS is safe and noninferior to the pre-specified performance goal in capturing debris in the vast majority of patients and with the use of a broad range of atherectomy systems.

Для цитирования: Shammas Nicolas W, Pucillo Anthony, Jenkins J Stephen, Garcia Lawrence A, Davis Tom, Aronow Herbert D, Armstrong Ehrin J, Zeller Thomas, Scheinert Dierk, Rosenschein Uri, Gray William. WIRION Embolic Protection System in Lower Extremity Arterial Interventions: Results of the Pivotal WISE LE Trial.. JACC. Cardiovascular interventions. 2018. DOI: 10.1016/j.jcin.2018.05.025.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.