Comparison of Visual Field Test Measurements With a Novel Approach on a Wearable Headset to Standard Automated Perimetry.

Johnson Catherine, Sayed Ahmed, McSoley John, Durbin Mary, Kashem Rashed, Nicklin Alexandra, Lopez Valeria, Mijares Georgeana, Chen Michael, Shaheen Abdulla, Segarra Steven, Rady Nadine, Duque Christian Andres, Opoku-Baah Collins, Abou Shousha Mohamed

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PRÉCIS: This study of inter-test comparability of a novel visual field application installed on an augmented-reality portable headset and Humphrey field analyzer Swedish interactive thresholding algorithm (SITA) Standard visual field test demonstrates the excellent correlation of mean deviation (MD) and mean sensitivity (MS).

PURPOSE: To determine the correlation between visual field testing with novel software on a wearable headset versus standard automated perimetry.

PATIENTS AND

METHODS

Patients with and without visual field defects attributable to glaucoma had visual field testing in one eye of each patient with 2 methods: re:Imagine Strategy (Heru, Inc.) and the Humphrey field analyzer (Carl Zeiss Meditec, Inc.) SITA Standard 24-2 program. Main outcome measures included MS and MD, which were evaluated by linear regression, intraclass correlation coefficient (ICC), and Bland Altman analysis for assessment of the mean difference and limits of agreement.

RESULTS

Measurements from 89 eyes of 89 patients (18 normal and 71 glaucomas) were compared with both instruments. Linear regression analysis demonstrated an excellent Pearson correlation coefficient of r = 0.94 for MS and r = 0.95 for MD. ICC analysis demonstrated high levels of concordance (ICC = 0.95, P < 0.001 for MS and ICC = 0.94, P < 0.001 for MD). Bland-Altman analysis determined a small mean difference between the two devices (Heru minus Humphrey) of 1.15 dB for MS and 1.06 dB for MD.

CONCLUSIONS

The Heru visual field test correlated well with SITA Standard in a population of normal eyes and eyes with glaucoma.

Для цитирования: Johnson Catherine, Sayed Ahmed, McSoley John, Durbin Mary, Kashem Rashed, Nicklin Alexandra, Lopez Valeria, Mijares Georgeana, Chen Michael, Shaheen Abdulla, Segarra Steven, Rady Nadine, Duque Christian Andres, Opoku-Baah Collins, Abou Shousha Mohamed. Comparison of Visual Field Test Measurements With a Novel Approach on a Wearable Headset to Standard Automated Perimetry.. Journal of glaucoma. 2023. DOI: 10.1097/ijg.0000000000002238.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.