Luithardt Annette Franziska, Meisner Christoph, Monhart Matthias, Krapp Elke, Mast Andrea, Schiefer Ulrich

СтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод в очереди

Аннотация

AIMS: To validate the EyeSuite version of German Adaptive Threshold Estimation (GATE), a new thresholding algorithm for automated static perimetry.

METHODS

Specification of agreement and its clinical evaluation as validation criteria. Comparison of local differential luminance sensitivity (DLS) and test time values between the prototype version of GATE (GATEp) and a clinical trial version, implemented in EyeSuite (GATEe), by means of modified Bland-Altman plots. All examinations were performed on the Octopus 900 perimeter (Haag-Streit Inc., Köniz, Switzerland). Visually impaired patients (anterior ischaemic optic neuropathy [n=3], glaucomatous optic neuropathy [n=15], (post-)chiasmal visual pathway lesion [n=6], retinitis pigmentosa [n=6]) were either tested with grid 30A (30° excentricity, 83 test locations) or grid 84NO (90° excentricity, 109 test locations, patients with RP only).

RESULTS

The comparison of local DLS values showed good-to-acceptable agreement between GATEp and GATEe (bias <2 dB, limits of agreement [LOA] <5 dB) and very good repeatability for GATEp (bias <0.5 dB, LOA<3 dB). Median test times for GATEp and GATEe were 7.8  and 8.8 min for the 30° grid and 6.7  and 7.8 min for the 90° grid.

CONCLUSIONS

GATEp and GATE, implemented in the commercially available EyeSuite software package (GATEe), show good agreement regarding local differential luminance sensitivity. GATEe can thus be also recommended for clinical practice.

CLINICAL TRIAL NUMBER: NCT01265628.

Для цитирования: Luithardt Annette Franziska, Meisner Christoph, Monhart Matthias, Krapp Elke, Mast Andrea, Schiefer Ulrich. Validation of a new static perimetric thresholding strategy (GATE).. The British journal of ophthalmology. 2014. DOI: 10.1136/bjophthalmol-2013-304535.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.