Phase Variance Analysis for Identifying Heterogeneous Conduction Regions in Persistent Atrial Fibrillation: Impact of Recording Duration.

Seno Hiroshi, Wang Yilin, Kojima Toshiya, Oshima Tsukasa, Yamagata Kenichiro, Yamazaki Masatoshi, Sakuma Ichiro, Fujiu Katsuhito, Tomii Naoki

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Phase variance analysis, which quantifies the spatial heterogeneity of local activation, has been proposed as a method to characterize complex conduction patterns in atrial fibrillation (AF). However, the influence of recording duration on its reliability remains unclear. This study aimed to determine how recording duration affects the accuracy and reproducibility of identifying heterogeneous conduction regions by phase variance analysis.

METHODS

Intracardiac electrogram recordings (30 s) were obtained from 16 left atrial sites in 5 patients with persistent AF. A reference phase variance map (REF) was generated by averaging instantaneous phase variance maps over the full 30-s dataset to represent the frequency of local conduction heterogeneity. Thirteen recording durations (0.01-25 s) were defined as experimental conditions. For each duration, 100 randomly selected segments were extracted, and estimated phase variance maps (ESTs) were computed using the same temporal averaging method. Accuracy was evaluated by calculating the structural similarity index measure (SSIM) between ESTs and REF, while reproducibility was assessed using pixel-wise standard deviations across subsamplings.

RESULTS

Longer recording durations produced higher SSIM values and lower standard deviations. With durations ≥ 10 s, SSIM consistently exceeded 0.95, including outliers, and standard deviations fell below 0.015, indicating accurate and stable estimation of complex conduction patterns.

CONCLUSIONS

Adequate recording duration enables accurate and reproducible estimation of heterogeneous conduction regions using phase variance analysis, even in persistent AF with complex conduction patterns. This method may support optimal ablation target selection and improve treatment outcomes.

Для цитирования: Seno Hiroshi, Wang Yilin, Kojima Toshiya, Oshima Tsukasa, Yamagata Kenichiro, Yamazaki Masatoshi, Sakuma Ichiro, Fujiu Katsuhito, Tomii Naoki. Phase Variance Analysis for Identifying Heterogeneous Conduction Regions in Persistent Atrial Fibrillation: Impact of Recording Duration.. Journal of arrhythmia. 2025. DOI: 10.1002/joa3.70246.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.