Inflammatory and Myocardial Biomarker Response Following Pulmonary Vein Isolation: Cryoballoon Versus Balloon-in-Basket Pulsed Field Ablation.

Wenzel Jan-Per, Hatahet Sascha, Eitel Charlotte, Abdessadok Raed, Popescu Sorin, Nikorowitsch Julius, de Waha Suzanne, Zeller Tanja, Kuck Karl-Heinz, Tilz Roland Richard

СтатьяСравнительное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Cryoballoon ablation (CB) is a well-established thermal technique for pulmonary vein isolation (PVI), while balloon-in-basket pulsed field ablation (BiB-PFA) represents a novel non-thermal alternative. Both single-shot systems may trigger systemic inflammation and myocardial injury, yet direct comparisons are lacking. This study aimed to compare inflammatory and myocardial biomarker responses following first-time PVI using CB or BiB-PFA.

METHODS

In this prospective, single-center study, 100 patients undergoing PVI for symptomatic paroxysmal or persistent atrial fibrillation (AF) were enrolled (CB: n =  50; BiB-PFA: n =  50). Venous blood samples were collected before and on the morning after ablation to assess leukocytes, C-reactive protein (CRP), platelets, troponin T, creatine kinase (CK), myoglobin, creatinine, and estimated glomerular filtration rate. Baseline characteristics, procedural data, and acute success were analyzed.

RESULTS

Patients in the CB group were older (74 vs. 65 years; p =  0.01) and had higher CHA₂DS₂-VAcore (3.0 vs. 2.0; p  =  0.009). Acute PVI was achieved in all cases CB was associated with greater increases in leukocytes (Δ2.5 vs. 1.1 × 10⁹/L; p  =  0.05) and CRP (Δ5.8 vs. 3.4 mg/L; p =  0.02), whereas BiB-PFA showed higher rises in CK (Δ217 vs. 103 U/L; p  =  0.01) and troponin T (Δ1129 vs. 614.5 ng/L; p =  0.01). No significant correlation was found between energy delivery and biomarker changes.

CONCLUSION

CB and BiB-PFA elicit distinct systemic responses. CB provoked stronger inflammatory activation, while BiB-PFA caused greater myocardial biomarker release, suggesting energy- and device-specific effects.

Для цитирования: Wenzel Jan-Per, Hatahet Sascha, Eitel Charlotte, Abdessadok Raed, Popescu Sorin, Nikorowitsch Julius, de Waha Suzanne, Zeller Tanja, Kuck Karl-Heinz, Tilz Roland Richard. Inflammatory and Myocardial Biomarker Response Following Pulmonary Vein Isolation: Cryoballoon Versus Balloon-in-Basket Pulsed Field Ablation.. Journal of cardiovascular electrophysiology. 2025. DOI: 10.1111/jce.70221.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.