Alterra Adaptive Prestent and SAPIEN 3 THV for Congenital Pulmonic Valve Dysfunction: An Early Feasibility Study.

Shahanavaz Shabana, Balzer David, Babaliaros Vasilis, Kim Dennis, Dimas Vivian, Veeram Reddy Suredranath R, Leipsic Jonathan, Blanke Philipp, Shirali Girish, Parthiban Anitha, Gorelick Jeremy, Zahn Evan M

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

The aim of this study was to demonstrate the safety and functionality of the Alterra Adaptive Prestent and SAPIEN 3 transcatheter heart valve (THV) in patients with dysfunctional, dilated right ventricular outflow tract (RVOT) greater or equal to moderate pulmonary regurgitation (PR).

BACKGROUND

Significant variations in the size and morphology of the RVOT affect the placement of transcatheter pulmonary valves. The Alterra Prestent internally reduces and reconfigures the RVOT, providing a stable landing zone for the 29-mm SAPIEN 3 THV.

METHODS

Eligible patients had moderate or greater PR, weighed >20 kg, and had RVOT diameter 27 to 38 mm and length >35 mm. The primary endpoint was device success, a 5-item composite: 1 Alterra Prestent deployed in the desired location, 1 SAPIEN 3 THV implanted in the desired location within the Prestent, right ventricular-to-pulmonary artery peak-to-peak gradient <35 mm Hg after THV implantation, less than moderate PR at discharge, and no explantation 24 h post-implantation. The secondary composite endpoint was freedom from THV dysfunction (RVOT/pulmonary valve (PV) reintervention, greater or equal to moderate total PR, mean RVOT/PV gradient ≥ 35 mm Hg at 30 days and 6 months. Descriptive statistics are reported.

RESULTS

Enrolled patients (N = 15) had a median age and weight of 20 years and 61.7 kg, respectively; 93.3% were in New York Heart Association functional class I or II. Device success was 100%. No staged procedures were necessary. No THV dysfunction was reported to 6 months. No serious safety signals were reported.

CONCLUSIONS

This early feasibility study demonstrated the safety and functionality of the Alterra Adaptive Prestent in patients with congenital RVOT dysfunction and moderate or greater PR. Durability and long-term outcome data are needed.

Для цитирования: Shahanavaz Shabana, Balzer David, Babaliaros Vasilis, Kim Dennis, Dimas Vivian, Veeram Reddy Suredranath R, Leipsic Jonathan, Blanke Philipp, Shirali Girish, Parthiban Anitha, Gorelick Jeremy, Zahn Evan M. Alterra Adaptive Prestent and SAPIEN 3 THV for Congenital Pulmonic Valve Dysfunction: An Early Feasibility Study.. JACC. Cardiovascular interventions. 2020. DOI: 10.1016/j.jcin.2020.06.039.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.