OBJECTIVES

Lung-protective mechanical ventilation aims to prevent alveolar collapse and overdistension, but reliable bedside methods to quantify them are lacking. We propose a quantitative descriptor of the shape of local pressure-volume curves derived from electrical impedance tomography, for computing maps that highlight the presence and location of regions of presumed tidal recruitment (i.e., elastance decrease during inflation, pressure-volume curve with upward curvature) or overdistension (i.e., elastance increase during inflation, downward curvature).

DESIGN

Secondary analysis of experimental cohort study.

SETTING

University research facility.

SUBJECTS: Twelve mechanically ventilated pigs.

INTERVENTIONS: After induction of acute respiratory distress syndrome by hydrochloric acid instillation, animals underwent a decremental positive end-expiratory pressure titration (steps of 2 cm H2O starting from ≥ 26 cm H2O).

MEASUREMENTS AND MAIN

RESULTS

Electrical impedance tomography-derived maps were computed at each positive end-expiratory pressure-titration step, and whole-lung CT taken every second steps. Airway flow and pressure were recorded to compute driving pressure and elastance. Significant correlations between electrical impedance tomography-derived maps and positive end-expiratory pressure indicate that, expectedly, tidal recruitment increases in dependent regions with decreasing positive end-expiratory pressure (p < 0.001) and suggest that overdistension increases both at high and low positive end-expiratory pressures in nondependent regions (p < 0.027), supporting the idea of two different scenarios of overdistension occurrence. Significant correlations with CT measurements were observed: electrical impedance tomography-derived tidal recruitment with poorly aerated regions (r = 0.43; p < 0.001); electrical impedance tomography-derived overdistension with nonaerated regions at lower positive end-expiratory pressures and with hyperaerated regions at higher positive end-expiratory pressures (r ≥ 0.72; p < 0.003). Even for positive end-expiratory pressure levels minimizing global elastance and driving pressure, electrical impedance tomography-derived maps showed nonnegligible regions of presumed overdistension and tidal recruitment.

CONCLUSIONS

Electrical impedance tomography-derived maps of pressure-volume curve shapes allow to detect regions in which elastance changes during inflation. This could promote individualized mechanical ventilation by minimizing the probability of local tidal recruitment and/or overdistension. Electrical impedance tomography-derived maps might become clinically feasible and relevant, being simpler than currently available alternative approaches.

Для цитирования: Beda Alessandro, Carvalho Alysson R, Carvalho Nadja C, Hammermüller Sören, Amato Marcelo B P, Muders Thomas, Gittel Claudia, Noreikat Katharina, Wrigge Hermann, Reske Andreas W. Mapping Regional Differences of Local Pressure-Volume Curves With Electrical Impedance Tomography.. Critical care medicine. 2017. DOI: 10.1097/ccm.0000000000002233.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.