MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/00034894211023218

Медицинская библиотека · The Annals of otology, rhinology, and laryngology · 2021

Selective Facial Muscle Activation with Acute and Chronic Multichannel Cuff Electrode Implantation in a Feline Model.

Sahyouni Ronald, Goshtasbi Khodayar, Presacco Alessandro, Birkenbeuel Jack, Cheung Dillon, Abiri Arash, Berger Michael H, Djalilian Hamid R, Lin Harrison W

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

Facial paralysis is a debilitating condition with substantial functional and psychological consequences. This feline-model study evaluates whether facial muscles can be selectively activated in acute and chronic implantation of 16-channel multichannel cuff electrodes (MCE).

METHODS

Two cats underwent acute terminal MCE implantation experiments, 2 underwent chronic MCE implantation in uninjured facial nerves (FN) and tested for 6 months, and 2 underwent chronic MCE implantation experiments after FN transection injury and tested for 3 months. The MCE were wrapped around the main trunk of the skeletonized FN, and data collection consisted of EMG thresholds, amplitudes, and selectivity of muscle activation.

RESULTS

In acute experimentation, activation of specific channels (ie, channels 1-3 and 6-8) resulted in selective activation of orbicularis oculi, whereas activation of other channels (ie, channels 4, 5, or 8) led to selective activation of levator auris longus with higher EMG amplitudes. MCE implantation yielded stable and selective facial muscle activation EMG thresholds and amplitudes up to a 5-month period. Modest selective muscle activation was furthermore obtained after a complete transection-reapproximating nerve injury after a 3-month recovery period and implantation reoperation. Chronic implantation of MCE did not lead to fibrosis on histology. Field steering was achieved to activate distinct facial muscles by sending simultaneous subthreshold currents to multiple channels, thus theoretically protecting against nerve damage from chronic electrical stimulation.

CONCLUSION

Our proof-of-concept results show the ability of an MCE, supplemented with field steering, to provide a degree of selective facial muscle stimulation in a feline model, even following nerve regeneration after FN injury.

LEVEL OF EVIDENCE: N/A.

Для цитирования: Sahyouni Ronald, Goshtasbi Khodayar, Presacco Alessandro, Birkenbeuel Jack, Cheung Dillon, Abiri Arash, Berger Michael H, Djalilian Hamid R, Lin Harrison W. Selective Facial Muscle Activation with Acute and Chronic Multichannel Cuff Electrode Implantation in a Feline Model.. The Annals of otology, rhinology, and laryngology. 2021. DOI: 10.1177/00034894211023218.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.