DOI 10.1007/s00330-019-06234-4

Медицинская библиотека · European radiology · 2019

High sensitivity and specificity of 4D-CTA in the detection of cranial arteriovenous shunts.

In 't Veld Matthijs, Fronczek Rolf, Dos Santos Marlise P, van Walderveen Marianne A A, Meijer Frederick J A, Willems Peter W A

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПеревод в очереди

Аннотация

PURPOSE: In a prospective cohort study, we evaluated the diagnostic accuracy of time-resolved CT angiography (4D-CTA) compared to digital subtraction angiography (DSA) for detecting cranial arteriovenous shunts.

MATERIAL AND

METHODS

Patients were enrolled if a DSA had been ordered querying either a dural arteriovenous fistula (dAVF) or a cerebral arteriovenous malformation (bAVM). After enrolment, both a DSA and a 4D-CTA were performed. Both studies were evaluated using a standardized form. If a dAVF or bAVM was found, its classification, angioarchitectural details, and treatment options were recorded.

RESULTS

Ninety-eight patients were enrolled and 76 full datasets were acquired. DSA demonstrated a shunting lesion in 28 out of 76 cases (prevalence 37%). 4D-CTA demonstrated all but two of these lesions (sensitivity of 93%) and produced one false positive (specificity of 98%). These numbers yielded a positive predictive value (PPV) of 96% and a negative predictive value (NPV) of 96%. Significant doubt regarding the 4D-CTA diagnosis was reported in 6.6% of all cases and both false-negative 4D-CTA results were characterized by such doubt.

CONCLUSIONS

4D-CTA has very high sensitivity and specificity for the detection of intracranial arteriovenous shunts. Based on these results, 4D-CTA may replace DSA imaging as a first modality in the diagnostic workup in a large number of patients suspected of a cranial dAVF or bAVM, especially if there is no doubt regarding the 4D-CTA diagnosis.

KEY POINTS: • 4D-CTA was shown to have a high diagnostic accuracy and is an appropriate, less invasive replacement for DSA as a diagnostic tool for cranial arteriovenous shunts in the majority of suspected cases. • Doubt regarding the 4D-CTA result should prompt additional DSA imaging, as it is associated with false negatives. • False-positive 4D-CTA results are rare, but do exist.

Для цитирования: In 't Veld Matthijs, Fronczek Rolf, Dos Santos Marlise P, van Walderveen Marianne A A, Meijer Frederick J A, Willems Peter W A. High sensitivity and specificity of 4D-CTA in the detection of cranial arteriovenous shunts.. European radiology. 2019. DOI: 10.1007/s00330-019-06234-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.