DOI 10.1016/j.jdent.2026.106858
Effect of a stackable scanning guide, intraoral scanner type, and operator on the accuracy of maxillary full-arch implant scans: An in vitro study.
Аннотация
PURPOSE: To evaluate how a stackable scanning guide, scanner type, and operator influence the accuracy of maxillary full-arch implant scans.
METHODS
A maxillary edentulous model with four implants was digitized with an industrial scanner (reference model) and scanned using 2 intraoral scanners (Primescan (PS) and Trios 4 (T4)), without (NG) and with a stackable scanning guide (SG), by two independent operators (n = 7). In addition, a conventional open-tray impressions (CNV) were made. Accuracy was assessed superimposing and assessing the 3D, angular, and interimplant distance deviations.
RESULTS
Impression type affected all 3D, angular, and 2D distance deviation outcomes, while operator influence was minimal (ICC 0-4%). CNV showed the greatest deviations. The scanning guide improved 3D accuracy in a position-dependent manner, with the greatest benefits observed at specific implant sites, particularly at distal positions, with deviations ranked CNV > T4 > PS. PS and T4 performed similarly without the guide, and having the guide a limited influence on angular and interimplant distance outcomes.
CONCLUSIONS
Digital implant scans were more accurate than conventional open-tray impressions for maxillary full-arch implant scans. The use of a stackable scanning guide improved scan trueness in a position-dependent manner, with scanner type influencing deviation patterns; Trios 4 demonstrated more consistent accuracy across implant positions, operator influence was minimal.
CLINICAL SIGNIFICANCE: A stackable scanning guide can improve full-arch digital implant scan accuracy by providing stable reference landmarks and potentially reducing stitching errors, particularly at distal implant sites.