MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jdent.2026.106858

Медицинская библиотека · Journal of dentistry · 2026

Effect of a stackable scanning guide, intraoral scanner type, and operator on the accuracy of maxillary full-arch implant scans: An in vitro study.

Cakmak Gülce, Bazelli Nando, Schenk Vanadis, Güven Mehmet Esad, Molinero-Mourelle Pedro, Abou-Ayash Samir

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate how a stackable scanning guide, scanner type, and operator influence the accuracy of maxillary full-arch implant scans.

METHODS

A maxillary edentulous model with four implants was digitized with an industrial scanner (reference model) and scanned using 2 intraoral scanners (Primescan (PS) and Trios 4 (T4)), without (NG) and with a stackable scanning guide (SG), by two independent operators (n = 7). In addition, a conventional open-tray impressions (CNV) were made. Accuracy was assessed superimposing and assessing the 3D, angular, and interimplant distance deviations.

RESULTS

Impression type affected all 3D, angular, and 2D distance deviation outcomes, while operator influence was minimal (ICC 0-4%). CNV showed the greatest deviations. The scanning guide improved 3D accuracy in a position-dependent manner, with the greatest benefits observed at specific implant sites, particularly at distal positions, with deviations ranked CNV > T4 > PS. PS and T4 performed similarly without the guide, and having the guide a limited influence on angular and interimplant distance outcomes.

CONCLUSIONS

Digital implant scans were more accurate than conventional open-tray impressions for maxillary full-arch implant scans. The use of a stackable scanning guide improved scan trueness in a position-dependent manner, with scanner type influencing deviation patterns; Trios 4 demonstrated more consistent accuracy across implant positions, operator influence was minimal.

CLINICAL SIGNIFICANCE: A stackable scanning guide can improve full-arch digital implant scan accuracy by providing stable reference landmarks and potentially reducing stitching errors, particularly at distal implant sites.

Для цитирования: Cakmak Gülce, Bazelli Nando, Schenk Vanadis, Güven Mehmet Esad, Molinero-Mourelle Pedro, Abou-Ayash Samir. Effect of a stackable scanning guide, intraoral scanner type, and operator on the accuracy of maxillary full-arch implant scans: An in vitro study.. Journal of dentistry. 2026. DOI: 10.1016/j.jdent.2026.106858.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.