JinJiang ablation catheter is no inferior than Johnson ablation catheter in radiofrequency ablation of paroxysmal supraventricular tachycardia.

Li Ying, Xiong Ganwei, Liu Yuan, Huang Xiantao, Liu Yanfeng, Zhou Haiwen, Xu Zhicheng, Tao Yu, Zhang Fanzhi, Lai Hengli, Chen Zhenhuan

Рандомизированное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Paroxysmal supraventricular tachycardia (PSVT) is characterized by episodes of rapid tachycardia with sudden onset and sudden termination. PSVT treatment has evolved considerably over the past 30 years. Currently, radiofrequency catheter ablation is the first-line treatment.

HYPOTHESIS: We conducted a randomized controlled trial to compare safety and effectiveness of PSVT ablation between the Jinjiang and Johnson (J&J) catheters in 57 patients in our hospital.

METHODS AND

RESULTS

Patients were randomly assigned to ablation procedures using either the Jinjiang system or the J&J Carto system. Follow-up was performed 3 days, 1, and 6 months after the procedure. Success rate, ablation time, frequency of ablation, and rates of complications and recurrence did not significantly differ between the groups. One Jinjiang group patient (3.6%) experienced arrhythmia recurrence during the 6-month follow-up.

CONCLUSIONS

The Jinjiang catheter for radiofrequency ablation of PSVT is as safe and effective as the J&J catheter.

Для цитирования: Li Ying, Xiong Ganwei, Liu Yuan, Huang Xiantao, Liu Yanfeng, Zhou Haiwen, Xu Zhicheng, Tao Yu, Zhang Fanzhi, Lai Hengli, Chen Zhenhuan. JinJiang ablation catheter is no inferior than Johnson ablation catheter in radiofrequency ablation of paroxysmal supraventricular tachycardia.. Pacing and clinical electrophysiology : PACE. 2023. DOI: 10.1111/pace.14898.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.