Incidence, predictors, and clinical impact of hypoattenuating leaflet thickening following SAPIEN 3 Ultra RESILIA implantation.

Ishizu Kenichi, Shirai Shinichi, Hayashi Masaomi, Morofuji Toru, Isotani Akihiro, Ohno Nobuhisa, Kakumoto Shinichi, Ando Kenji, Yamamoto Masanori, Ochiai Tomoki, Tsunaki Tatsuya, Hioki Hirofumi, Shimura Tetsuro, Yashima Fumiaki, Asami Masahiko, Yamanaka Futoshi, Ohno Yohei, Nakazawa Gaku, Hachinohe Daisuke, Fuku Yasushi, Otsuka Toshiaki, Hayashida Kentaro, Of The Ocean-Tavi Investigators On Behalf

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

The latest-generation SAPIEN 3 Ultra RESILIA (S3UR) transcatheter heart valve (THV) incorporates several changes in leaflet design, including an improved anticalcification coating and modified commissural attachment. There are no established data on hypoattenuating leaflet thickening (HALT) following transcatheter aortic valve implantation (TAVI) using the S3UR.

AIMS: Our study aimed to elucidate the clinical features of HALT following S3UR implantation.

METHODS

As a subset of the OCEAN (Optimized CathEter vAlvular INtervention)-TAVI registry, we prospectively assessed patients who underwent cardiac computed tomography (CT) 30 days after S3UR implantation. HALT and potentially relevant THV geometry were analysed using four-dimensional CT data by an independent core laboratory.

RESULTS

Of the 445 patients studied, HALT was detected in 95 patients (21.3%) 30 days after TAVI. The modification of the commissural attachment specific to the 20 mm and 23 mm S3UR THVs did not affect the incidence of HALT (22.1% for ≤23 mm; 20.2% for ≥26 mm; p=0.636). The hourglass-shaped THV frame (p<0.001) and asymmetricity of THV leaflets (p=0.002) were independently associated with HALT development. A trend toward higher mean aortic gradients at 30 days with greater degrees of HALT (HALT >25% vs HALT ≤25%: 10.3 [interquartile range [IQR] 7.0-13.0] mmHg vs 8.6 [IQR 6.3-11.6] mmHg; p=0.007; HALT >50% vs HALT ≤50%: 11.5 [IQR 7.0-14.3] mmHg vs 8.9 [IQR 6.3-11.9] mmHg; p=0.002) was noted.

CONCLUSIONS

The incidence of HALT for the S3UR was comparable with the already reported incidences for the previous-generation SAPIEN 3 THV. Given the haemodynamic impact of HALT severity and multiplicity, strategic planning to avoid deformation of the implanted THV might be required. (Clinical trial registration: UMIN000020423).

Для цитирования: Ishizu Kenichi, Shirai Shinichi, Hayashi Masaomi, Morofuji Toru, Isotani Akihiro, Ohno Nobuhisa, Kakumoto Shinichi, Ando Kenji, Yamamoto Masanori, Ochiai Tomoki, Tsunaki Tatsuya, Hioki Hirofumi, Shimura Tetsuro, Yashima Fumiaki, Asami Masahiko, Yamanaka Futoshi, Ohno Yohei, Nakazawa Gaku, Hachinohe Daisuke, Fuku Yasushi, Otsuka Toshiaki, Hayashida Kentaro, Of The Ocean-Tavi Investigators On Behalf. Incidence, predictors, and clinical impact of hypoattenuating leaflet thickening following SAPIEN 3 Ultra RESILIA implantation.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2025. DOI: 10.4244/eij-d-25-00523.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.