A Quantitative Method for the Measurement of Dried Blood Spot Amino Acids Using Ultra-Performance Liquid Chromatography.
Аннотация
BACKGROUND
Measurement of amino acids in dried blood spots has been extensively used for the detection of newborns with various inborn errors of amino acid metabolism including phenylketonuria (PKU) and maple syrup urine disease (MSUD). Whereas blood spot amino acid measurement has been invaluable for initial diagnosis, the relative insensitivity of blood spot measurement has found limited use in lifelong monitoring of patients with these disorders. The work described here outlines the evaluation of blood spot amino acid analysis using ultra-performance liquid chromatography (UPLC©) for use in follow-up testing.
METHODS
Dried blood spot amino acids were derivatized with a proprietary AccQTag® reagent and separated using UPLC. Plasma amino acids from dried bloods spots were obtained from 318 patient samples and compared to corresponding plasma samples measured using the same UPLC method.
RESULTS
Dried blood spot amino acid concentrations were highly correlated but negatively biased vs plasma concentrations. Interassay imprecision studies using UPLC demonstrated a %CV for phenylalanine of 4.81%-16.07%, tyrosine 5.62%-20.16%, valine 4.23%-15.46%, leucine 8.3%-15.3%, and isoleucine 4.25%-16.80%. Intraassay imprecision studies using UPLC demonstrated a %CV for phenylalanine of 0.42%-3.4%, tyrosine 1.6%-7.85%, valine 0.14%-1.84%, leucine 0.28%-2.01%, and isoleucine 0.6%-2.65%. Blood spot amino acid concentrations were stable for at least 3 days at temperatures up to 65 °C.
CONCLUSIONS
This UPLC-based method can reliably measure clinically significant amino acids in dried blood spots.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.