Early echocardiographic changes after percutaneous implantation of the Edwards SAPIEN transcatheter heart valve in the pulmonary position.

Chowdhury Shahryar M, Hijazi Ziyad M, Rhodes John, Kar Saibal, Makkar Raj, Mullen Michael, Cao Qi-Ling, King Lydia, Akin Jodi, Shirali Girish

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

To evaluate echocardiographic changes after SAPIEN valve implantation in the pulmonary position.

BACKGROUND

The feasibility of the SAPIEN transcatheter pulmonary valve (TPV) has recently been demonstrated. We evaluated changes in pulmonary valve function and the right ventricle after SAPIEN TPV placement.

METHODS

We evaluated echocardiograms at baseline, discharge, 1 and 6 months after TPV placement in 33 patients from 4 centers. Pulmonary insufficiency severity was graded 0-4. TPV peak and mean gradients were measured. Right ventricular (RV) size and function were quantified using routine measures derived from color, spectral, and tissue Doppler indices and two-dimensional echocardiography.

RESULTS

At baseline, 94% patients demonstrated pulmonary insufficiency grade 2-4. This decreased to 12% patients at 6 months (P < 0.01). TPV peak (P < 0.01) and mean gradient (P < 0.01) decreased. RV end-diastolic area indexed to body surface area (BSA) (P < 0.01), Tricuspid regurgitation (TR) gradient (P < 0.01), and the ratio of TR jet area to BSA (P < 0.01) decreased. Tricuspid inflow peak E:A, tissue Doppler imaging (TDI): septal E' and A', TDI: tricuspid A' improved between baseline and discharge, but trended back to baseline by 6-month follow-up. Tricuspid valve annulus z-score, RV area change, tricuspid annular plane systolic excursion (TAPSE), RV dP/dt, tricuspid E:E', and TDI: tricuspid annulus E' showed no change.

CONCLUSION

Improvements in pulmonary insufficiency and stenosis, RV size, and TR gradient and severity are seen after SAPIEN TPV placement. Selected indices of RV diastolic function improve immediately after TPV implantation, but return to baseline by 6 months. RV systolic function is unchanged.

Для цитирования: Chowdhury Shahryar M, Hijazi Ziyad M, Rhodes John, Kar Saibal, Makkar Raj, Mullen Michael, Cao Qi-Ling, King Lydia, Akin Jodi, Shirali Girish. Early echocardiographic changes after percutaneous implantation of the Edwards SAPIEN transcatheter heart valve in the pulmonary position.. Echocardiography (Mount Kisco, N.Y.). 2013. DOI: 10.1111/echo.12147.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.