Santoro Amato, Ziacchi Matteo, Calvanese Raimondo, Rapacciuolo Antonio, Volpicelli Mario, Poggio Luca, Nocerino Pasquale, Castagno Davide, Bianchi Valter, Magliano Paolo Liberato, Dell'Era Gabriele, Mandurino Cosimo, Mirizzi Gianluca, Savarese Gianluca, Marinaccio Leonardo, La Greca Carmelo, Covino Gregorio, Zingarini Gianluca, Varalda Marco, Vitali Francesco, Baiocchi Claudia, Campari Monica, Valsecchi Sergio, Bertini Matteo

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Left bundle branch area pacing (LBBAP) has emerged as a promising physiological pacing strategy. However, its feasibility and safety using conventional stylet-driven leads (SDL) such as INGEVITY+ remain under investigation.

OBJECTIVE

To evaluate the clinical performance, safety, and follow-up outcomes of LBBAP using the INGEVITY+ lead in a real-world electrophysiology setting.

METHODS

From 2021 to 2024, 207 consecutive patients underwent LBBAP implant attempts using the INGEVITY+ lead across 19 centers. Of these, 146 had bradycardia indications, and 61 had heart failure indications. A control group of 200 patients received standard right ventricular (RV) pacing with the same lead. Procedural outcomes, electrical parameters, complications, and mid-term lead performance were assessed.

RESULTS

LBBAP was successfully achieved in 201 (97%) cases. Procedural and fluoroscopy times were slightly longer for LBBAP than for RV pacing. Acute complications included one helix fracture, one atrioventricular block, and one septal perforation. During a median follow-up of 9 months, no lead fractures were reported; lead dislodgment occurred in three cases. Kaplan-Meier analysis showed no significant difference in time to first lead-related complication between the LBBAP and RV pacing groups (hazard ratio: 4.69, 95%CI: 0.63-34.90, p = 0.139). Electrical performance remained stable, with 98% of retained leads maintaining capture thresholds ≤ 2 V and 92% with sensed amplitudes ≥ 5 mV.

CONCLUSIONS

LBBAP using the INGEVITY + SDL is feasible and safe in clinical practice, with excellent implant success and stable mid-term electrical performance. These findings support its use as a viable conduction system pacing strategy.

Для цитирования: Santoro Amato, Ziacchi Matteo, Calvanese Raimondo, Rapacciuolo Antonio, Volpicelli Mario, Poggio Luca, Nocerino Pasquale, Castagno Davide, Bianchi Valter, Magliano Paolo Liberato, Dell'Era Gabriele, Mandurino Cosimo, Mirizzi Gianluca, Savarese Gianluca, Marinaccio Leonardo, La Greca Carmelo, Covino Gregorio, Zingarini Gianluca, Varalda Marco, Vitali Francesco, Baiocchi Claudia, Campari Monica, Valsecchi Sergio, Bertini Matteo. Implantation Success, Electrical Performance, and Safety of an Active Fixation Stylet-Driven Lead for LBBAP in Clinical Practice: A Multicenter Experience.. Journal of cardiovascular electrophysiology. 2025. DOI: 10.1111/jce.70237.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.