NEMESIS-PFA: Investigating Collateral Tissue Injury Associated With Pulsed Field Ablation.

Lakkireddy Dhanunjaya, Katapadi Aashish, Garg Jalaj, Herink Eli, Klotz Michael, Ganta Jagruth, Kabra Aanya, Kabra Rajesh, Pothineni Naga Venkata, Darden Douglas, Tummala Rangarao, Koerber Scott, Gopinathannair Rakesh, Bommana Sudharani, Atkins Donita, Chaudhary Rahul, Mbai Mackenzie, Tholakanahalli Venkat, Mohanty Sanghamitra, DiBiase Luigi, Natale Andrea

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Early evidence showed reduced complications with pulsed field ablation (PFA), but non-target tissue collateral damage created by electroporation effects is poorly understood and may significantly differ between systems.

OBJECTIVES

In this study, the authors evaluated the collateral effects of PFA.

METHODS

NEMESIS-PFA is a multicenter, observational registry of patients who underwent AF ablation from March 2024 onwards with any approved PFA systems-either a circular multielectrode array, spherical, pentaspline, or variable loop catheter-or radiofrequency ablation (RFA). We assessed procedural characteristics, biomarkers for myocardial injury, hemolytic anemia, and renal function, and left atrial function in select patients.

RESULTS

A total of 871 patients, aged 68.9 ± 10.9 years and male (70.8%), with paroxysmal atrial fibrillation (59.4%), and CHA2DS2VASC of 3.3 ± 1.3 were included. Of these, 87.1% (n = 773) underwent PFA with a pentaspline (70.9%), circular multielectrode (14.1%), spherical (12.4%), and variable loop (2.3%) catheter. Significant postprocedural change in certain biomarkers such as troponin (13,551.0 vs 127.5 ng/dL; P < 0.001), lactate dehydrogenase (107.5 vs 26.5 IU/L; P < 0.001), and haptoglobin (-102.0 vs -33.5 mg/dL; P < 0.001) were detected following the PFA procedures compared with RFA, and the change was dose-dependent. There were also significant differences in biomarkers across PFA systems. Lastly, there was a significant change in left atrial ejection fraction (-20.0% vs. -5.0%, P < 0.001) in PFA vs RFA.

CONCLUSIONS

Current PFA technologies are associated with worse troponin leak, hemolysis, and renal dysfunction than RFA. As PFA becomes mainstream, future studies appraising these effects and understanding the short term and long-term implications are needed.

Для цитирования: Lakkireddy Dhanunjaya, Katapadi Aashish, Garg Jalaj, Herink Eli, Klotz Michael, Ganta Jagruth, Kabra Aanya, Kabra Rajesh, Pothineni Naga Venkata, Darden Douglas, Tummala Rangarao, Koerber Scott, Gopinathannair Rakesh, Bommana Sudharani, Atkins Donita, Chaudhary Rahul, Mbai Mackenzie, Tholakanahalli Venkat, Mohanty Sanghamitra, DiBiase Luigi, Natale Andrea. NEMESIS-PFA: Investigating Collateral Tissue Injury Associated With Pulsed Field Ablation.. JACC. Clinical electrophysiology. 2025. DOI: 10.1016/j.jacep.2025.04.017.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.