Медицинская библиотека · Clinical oral implants research · 2026

Accuracy and Efficiency of Dynamic Computer-Assisted Implant Surgery Using Three Different Registration Methods-A Laboratory Study.

Nojiri Toshiki, Pala Kevser, Yamamoto Ryo, Fukutoku Akihiro, Kon Kazuhiro, Nagai Shigemi, Gallucci German O

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Accurate implant placement with Computer-Assisted Implant Surgery (CAIS) is critical to ensure long-term success. Dynamic-CAIS systems enhance surgical precision through real-time feedback, though comparative data on their accuracy and efficiency remain limited. This study evaluated how different registration methods in d-CAIS systems influence implant placement accuracy and procedural time.

MATERIALS AND

METHODS

In this laboratory study, three registration methods were assessed: ND group, XC group, and XM group. Five experienced operators placed 25 implants per group (75 total) in partially edentulous maxillary models. Postoperative CBCT scans were used to measure deviations at the implant entry point, apex, vertical depth, and angle. One-way ANOVA and Tukey's HSD test were used for statistical analysis.

RESULTS

Mean 3D deviations at the implant entry point were 1.06 mm (ND), 0.71 mm (XC), and 0.90 mm (XM); at the apex 1.23 mm, 0.83 mm, and 1.07 mm, respectively. Vertical depth deviations at the apex were similar across groups: 0.61 mm (ND), 0.59 mm (XC), and 0.64 mm (XM). Angular deviation was highest in the ND group (2.96°) compared to XC (1.17°) and XM (1.02°), with a significant difference between ND and XM (p = 0.028). The average procedural time was shortest in the XM group (7.20 ± 2.48 min), though differences were not statistically significant.

CONCLUSIONS

Although registration methods minimally affect overall accuracy, system-specific factors may influence angular deviation. Future studies should prioritize clinical trials with larger sample sizes and explore the influence of operator experience and learning curves.

Для цитирования: Nojiri Toshiki, Pala Kevser, Yamamoto Ryo, Fukutoku Akihiro, Kon Kazuhiro, Nagai Shigemi, Gallucci German O. Accuracy and Efficiency of Dynamic Computer-Assisted Implant Surgery Using Three Different Registration Methods-A Laboratory Study.. Clinical oral implants research. 2026. DOI: 10.1111/clr.70140.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.