DOI 10.1016/j.hrthm.2025.04.031

Медицинская библиотека · Heart rhythm · 2025

One-year outcomes of a conformable single-shot pulsed-field ablation catheter for the treatment of paroxysmal atrial fibrillation.

Reddy Vivek Y, Peichl Petr, Kautzner Josef, Anter Elad, Metzner Andreas, Koruth Jacob, Jais Pierre, Rackauskas Gediminas, Petru Jan, Funasako Moritoshi, Marinskis Germanas, Turagam Mohit, Aidietis Audrius, Selma Jada M, Nejedlo Vojtech, Kueffer Fred, Tarakji Khaldoun G, Natale Andrea, Neuzil Petr

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Most single-shot pulsed-field ablation (PFA) catheters require extensive repositioning for pulmonary vein isolation (PVI), posing a challenge for obtaining contiguous, durable lesions.

OBJECTIVE

To determine 1-year outcomes of a single-shot, all-in-one mapping and ablation PFA catheter for treating paroxysmal atrial fibrillation (PAF).

METHODS

After PVI with the large-lattice catheter with expandable tip (Sphere-360), follow-up included Holter monitoring at 180 and 365 days and scheduled/symptomatic trans-telephonic monitoring (TTM) or modeled insertable loop recorder (ILR) data. Efficacy outcomes were acute PVI and 12-month freedom from atrial arrhythmias (AA), after 90-day blanking. Optional invasive remapping at 75 days facilitated waveform refinement from PULSE1, PULSE2, to the optimized PULSE3.

RESULTS

At 3 centers, 100 PAF patients underwent PFA with PULSE1 (n = 30), PULSE2 (n = 20), or PULSE3 (n = 50). Procedure, left atrial dwell, and fluoroscopy times were 57.9 ± 20.6, 22.2 ± 11.8 and 6.8 ± 5.7 minutes, respectively. All 395 targeted PVs were acutely isolated, with a transpired PVI time of 11.5 ± 6.0 minutes, using 4.0 ± 1.3 lesions/PV. There were no primary safety events (serious device-related events within 7 days post-PFA). PVI durability with PULSE3 (n = 40) was 98% (per-vein) and 93% (per-patient). One-year freedom from AA recurrence was 82.0% (95% CI:73.0%-88.3%) overall, and 88.0% (95%CI, 75.2%-94.4%) for PULSE3 patients. Of the ILR sub-cohort (n = 15 PULSE3 patients), 3 patients (20%) had recurrences, with an AA burden reduction from 26% (baseline) to 1.6% (post-ablation).

CONCLUSION

The large lattice PFA catheter was efficient, safe, and effective in treating PAF. The observed high PVI durability translated to clinical effectiveness, even in continuously monitored patients.

Для цитирования: Reddy Vivek Y, Peichl Petr, Kautzner Josef, Anter Elad, Metzner Andreas, Koruth Jacob, Jais Pierre, Rackauskas Gediminas, Petru Jan, Funasako Moritoshi, Marinskis Germanas, Turagam Mohit, Aidietis Audrius, Selma Jada M, Nejedlo Vojtech, Kueffer Fred, Tarakji Khaldoun G, Natale Andrea, Neuzil Petr. One-year outcomes of a conformable single-shot pulsed-field ablation catheter for the treatment of paroxysmal atrial fibrillation.. Heart rhythm. 2025. DOI: 10.1016/j.hrthm.2025.04.031.