Real-World Safety of Two Pulsed Field Ablation Catheters for Atrial Fibrillation: A MAUDE Analysis.

Shoura Sami J, Singireddy Shreyas, Coombes Kyle, Saravanabavanandan Ragavendar, Murphy Skylar, Shilbayeh Abdul-Rahim, Bhardwaj Chetan, Shah Siddharth

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Two pulsed field ablation (PFA) catheter systems for atrial fibrillation (AF) are widely adopted: the Affera Sphere-9 (Medtronic) and the Varipulse (Biosense Webster). Their comparative real-world safety is unknown.

METHODS

We searched the FDA MAUDE database (June 27, 2026) for all reports naming either catheter since U.S. release. After excluding duplicates and non-catheter components, 743 Affera and 361 Varipulse reports remained. Reports were classified by FDA Event Type (device- vs. patient-related) and complication status, with complications grouped by organ system. Proportions were compared with chi-square or Fisher exact tests in R 4.3.3.

RESULTS

Neurologic events were far more frequent among Varipulse reports (44.3 % vs. 14.8%; P < 0.001), driven by stroke (38.0% vs. 9.7%; P < 0.001). Cardiovascular events predominated among Affera reports (43.2% vs. 29.1%; P < 0.001), including ventricular fibrillation (52 vs. 1), heart block (8.7% vs. 1.4%; P < 0.001), and hypotension (5.5% vs. 0.8%; P<0.001); effusion/tamponade was more frequent with Varipulse (15.2% vs. 9.0%; P = 0.003). Complication-free reports were more frequent with Affera (35.3% vs. 21.1%; P < 0.001). Deaths did not differ (2.6% vs 1.4%; P = 0.30).

CONCLUSIONS

These PFA systems show distinct reported safety signatures: neurologic/stroke-predominant for Varipulse and cardiovascular/arrhythmic (ventricular fibrillation, heart block) for Affera. Because MAUDE lacks procedural denominators, these proportions describe reported events, not incidence. Prospective, patient-level studies are needed to establish causation.

Для цитирования: Shoura Sami J, Singireddy Shreyas, Coombes Kyle, Saravanabavanandan Ragavendar, Murphy Skylar, Shilbayeh Abdul-Rahim, Bhardwaj Chetan, Shah Siddharth. Real-World Safety of Two Pulsed Field Ablation Catheters for Atrial Fibrillation: A MAUDE Analysis.. Pacing and clinical electrophysiology : PACE. 2026. DOI: 10.1111/pace.70396.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.