Syncope due to Pacing Inhibition During VT Detection in a Conduction System Pacing ICD Programmed in AAI(R) Mode.

García Rubén, Riesgo Álvaro, Tornero Raquel, Olaskoaga Leire, Martínez Carolina, Casas Maitane, Luque Óscar, Vázquez-Calvo Sara

Клинический случайСтатьяПеревод в очереди

Аннотация

The lack of dedicated generators for conduction system pacing can force the use of unconventional device configurations in selected patients. We report the case of a 74-year-old pacing-dependent man with non-ischemic cardiomyopathy who underwent an upgrade to LBBAP-ICD, programmed in AAI(R) mode. Syncope occurred after a nonsustained ventricular tachycardia episode due to activation of the Ventricular Tachycardia Response algorithm in a Boston Scientific ICD. This case highlights a potential interaction between device algorithms and pacing configuration that may result in clinically significant pauses in pacing-dependent patients.

Для цитирования: García Rubén, Riesgo Álvaro, Tornero Raquel, Olaskoaga Leire, Martínez Carolina, Casas Maitane, Luque Óscar, Vázquez-Calvo Sara. Syncope due to Pacing Inhibition During VT Detection in a Conduction System Pacing ICD Programmed in AAI(R) Mode.. Pacing and clinical electrophysiology : PACE. 2026. DOI: 10.1111/pace.70337.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.