MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11748-020-01415-1

Медицинская библиотека · General thoracic and cardiovascular surgery · 2020

Artificial endocrine pancreas with a closed-loop system effectively suppresses the accelerated hyperglycemic status after reperfusion during aortic surgery.

Aizawa Kei, Muraoka Arata, Kurumisawa Soki, Akutsu Hirohiko, Sugaya Akira, Uesugi Satoshi, Kawahito Koji

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To control intraoperative hyperglycemia in patients who underwent aortic surgery using STG-55® artificial endocrine pancreas and clarify the effectiveness of this device.

METHODS

Blood glucose control using the STG-55® was performed in 18 patients (15 men and 3 women; age, 66 ± 10 years) who required hypothermic circulatory arrest (STG-55® group). Seventeen patients (10 men and 7 women; age, 71 ± 8 years) whose blood glucose was controlled using the conventional method were included in the control group. Glucose concentration was controlled with the aim of maintaining it at 150 mg/dl.

RESULTS

In both groups, the blood glucose concentrations did not significantly change during the interruption of systemic perfusion; however, a sharp increase was noted immediately after reperfusion. Although the hyperglycemic status persisted after reperfusion in the control group, it was effectively suppressed in the STG-55® group (STG® vs. control group at 50 min after reperfusion: 180 ± 35 vs. 212 ± 47 mg/dl, p = 0.026) and blood glucose concentration reached the target value of 150 mg/dl at 100 min after reperfusion (STG® vs. control group: 153 ± 29 vs. 215 ± 43 mg/dl, p = 0.0008). The total administered insulin dose was 175 ± 81 U and 5 ± 3 U in the STG® and control groups, respectively (p < 0.0001).

CONCLUSIONS

To treat the accelerated hyperglycemic status in aortic surgery requiring circulatory arrest, strict glycemic control using an artificial endocrine pancreas might be beneficial.

Для цитирования: Aizawa Kei, Muraoka Arata, Kurumisawa Soki, Akutsu Hirohiko, Sugaya Akira, Uesugi Satoshi, Kawahito Koji. Artificial endocrine pancreas with a closed-loop system effectively suppresses the accelerated hyperglycemic status after reperfusion during aortic surgery.. General thoracic and cardiovascular surgery. 2020. DOI: 10.1007/s11748-020-01415-1.