SDFStent: Real-time interactive virtual stenting via SDF deformation fields.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Li Bohan J, Dorn Nicholas C, Lasso Andras, Jolley Matthew A, Feinstein Jeffrey A, James Doug L, Marsden Alison L

Аннотация

BACKGROUND AND OBJECTIVE: Stenting is among the most common transcatheter interventions for congenital heart disease (CHD). Patient-specific computational fluid dynamics (CFD) simulations can predict hemodynamic outcomes of intervention scenarios but require post-operative vascular geometries that reflect stent-induced shape changes, which existing tools either model inadequately or require extensive time or manual effort to generate. We present SDFStent, a signed distance function (SDF) based mesh deformation method for virtual stenting that operates in real time, maintains mesh integrity, and preserves junction geometry.

METHODS: The stent is modeled as a pipe surface composed of piecewise-capsule SDFs joined by a smooth-minimum operator. Mesh vertices near the expanding SDF surface are displaced along the SDF gradient with a compactly supported fall-off function and an alpha blending mask. SDFStent was benchmarked against three existing approaches and validated on a proof-of-concept cohort of three tetralogy of Fallot (ToF) and three coarctation of the aorta (CoA) patients using rigid-wall steady-state CFD simulations against clinical catheterization measurements.

RESULTS: Against a prescribed diameter of 6.0 mm, the method produced a mean stented diameter of 5.92 ± 0.08 mm in 1.5 s, over 100× faster than the best stenting-specific comparator. All output meshes were watertight and self-intersection-free. CFD-simulated post-operative pressure drops agreed with clinical measurements within 4 mmHg (mean error 2 mmHg).

CONCLUSIONS: SDFStent produces simulation-ready post-stent models that match prescribed stent dimensions at interactive speeds, from pre-operative anatomy and catheterization data alone. The implementation is open-source and available in 3D Slicer. Its scriptable architecture enables automated generation of large synthetic cohorts for data-driven surrogate modeling.