One-Year Outcomes of Novel Transcatheter Edge-to-Edge Mitral Repair System in Patients With Functional Mitral Regurgitation.

Xu Kai, Zhu Da, Jiang Hong, Chen Jing, Chen Shaoliang, Zhang Junjie, He Ben, Wang Yan, Fu Guosheng, Chen Jiyan, An Jian, Xiu Jiancheng, Guo Xiaogang, Li Yue, Cheng Xiang, Li Ping, Chen Yuguo, Zhou Shenghua, Sun Yingxian, Yu Bo, Pan Xiangbin, Han Yaling

СтатьяМногоцентровое исследованиеКлиническое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Transcatheter edge-to-edge repair (TEER) improves outcomes in patients with heart failure and moderate-to-severe functional mitral regurgitation (FMR) who remain symptomatic despite the use of maximal doses of guideline-directed medical therapy (GDMT). The SQ-Kyrin-M system is a novel TEER device designed for FMR treatment.

AIMS: This multicenter, prospective, single-arm study (ClinicalTrials.gov number: NCT05988450) evaluates the safety and efficacy of the SQ-Kyrin-M system in severe symptomatic FMR.

METHODS

A total of 125 eligible patients (mean age 65.5 ± 8.1 years) with heart failure and moderate-to-severe or severe FMR who remained symptomatic despite the use of maximal doses of GDMT were involved in the analysis. The primary endpoint was the composite rate of all-cause mortality and heart failure hospitalization at 1 year post-intervention. Independent assessments were conducted by an echocardiography core laboratory (ECL) and a clinical events committee (CEC).

RESULTS

At 1 year, the primary endpoint rate was 16.0% (20/125), and the composite major adverse event rate was 10.4%. Mitral regurgitation was reduced to ≤ 2+ in 94.2% of patients, with significant left ventricular reverse remodeling (LVEDV reduction: 33.8 ± 55.4 mL, p < 0.001). Patient-reported outcomes improved significantly, with Kansas City Cardiomyopathy Questionnaire (KCCQ) scores increasing by 15.9 ± 18.8 points (p < 0.001) and 6-minute walk distance (6MWD) improving by 46.7 ± 90.7 m (p < 0.001). The proportion of patients in New York Heart Association (NYHA) class I/II increased from 30.4% to 84.4% (p < 0.001).

CONCLUSIONS

The SQ-Kyrin-M transcatheter mitral valve repair system bolsters the evidence supporting the safety and efficacy of TEER interventions for FMR patients.

Для цитирования: Xu Kai, Zhu Da, Jiang Hong, Chen Jing, Chen Shaoliang, Zhang Junjie, He Ben, Wang Yan, Fu Guosheng, Chen Jiyan, An Jian, Xiu Jiancheng, Guo Xiaogang, Li Yue, Cheng Xiang, Li Ping, Chen Yuguo, Zhou Shenghua, Sun Yingxian, Yu Bo, Pan Xiangbin, Han Yaling. One-Year Outcomes of Novel Transcatheter Edge-to-Edge Mitral Repair System in Patients With Functional Mitral Regurgitation.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2025. DOI: 10.1002/ccd.70230.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.