Recurrent inappropriate ICD shocks following lead fracture due to blunt chest Trauma: A case report.

Man Singh Anshu, Srivastava Shubham, Khanna Roopali, Kapoor Aditya

Клинический случайСтатьяПеревод в очереди

Аннотация

Implantable cardioverter-defibrillators (ICDs) are effective in preventing sudden cardiac death; however, lead-related complications remain an important cause of device malfunction. Lead fracture may result in oversensing and inappropriate shocks, significantly affecting patient morbidity. We report a 48-year-old man with ischemic cardiomyopathy and a single-chamber ICD who presented with multiple inappropriate shocks following a motor vehicle accident. Device interrogation demonstrated high-frequency non-physiological noise, abrupt reduction in lead impedance, and premature battery depletion. Fluoroscopy confirmed ventricular lead fracture. The fractured lead was abandoned, and a new ventricular lead was successfully implanted. Inappropriate shocks resolved completely following lead revision. Trauma-related lead fracture should be considered in patients presenting with inappropriate ICD therapies after blunt chest trauma. Prompt device interrogation and lead revision are essential for appropriate management.

Для цитирования: Man Singh Anshu, Srivastava Shubham, Khanna Roopali, Kapoor Aditya. Recurrent inappropriate ICD shocks following lead fracture due to blunt chest Trauma: A case report.. Indian pacing and electrophysiology journal. 2026. DOI: 10.1016/j.ipej.2026.09.008.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.