OBJECTIVES

Redo sternotomy and explantation of left ventricular assist devices (LVAD) for heart transplantation (HT) involve prolonged dissection, potential injury to mediastinal structures and/or bleeding. Our study compared a complete expanded polytetrafluoroethylene (ePTFE) wrap versus minimal or no ePTFE during LVAD implantation, on outcomes of subsequent HT.

METHODS

Between July 2005 and July 2018, 84 patients underwent a LVAD implant and later underwent HT. Thirty patients received a complete ePTFE wrap during LVAD implantation (Group 1), and 54 patients received either a sheet of ePTFE placed in the anterior mediastinum or no ePTFE (Group 2).

RESULTS

Baseline characteristics were similar between Groups 1 and 2. Surgeons reported subjective improvements in speed, predictability, and safety of dissection with complete ePTFE compared with minimal or no ePTFE. Time from incision to initiation of cardiopulmonary bypass (CPB) were similar between groups (97 ± 38 vs. 89 ± 29 min, p = .3). Injury to mediastinal structures during the dissection was similar between groups (10% vs. 11%, p > .9). While surgeons reported less intraoperative bleeding in Group 1 (43% vs. 61%), this trend did not reach significance (p = .1). In-hospital mortality, intensive care unit length of stay and hospital length of stay were similar between both groups.

CONCLUSIONS

In patients undergoing LVAD explant-HT, there was a trend toward reduced surgeon reported intraoperative bleeding with ePTFE placement. Despite qualitatively reported greater ease and speed of mediastinal dissection with ePTFE membrane placement, time to initiation of CPB did not differ, likely because surgeons remained cautious, allowing extra time for unanticipated difficulties.

Для цитирования: An Kevin R, Christakis Nicole, Jegatheeswaran Anusha, Cusimano Robert J, Rao Vivek, Badiwala Mitesh, Yau Terrence M. Outcomes of expanded polytetrafluoroethylene pericardial membrane implantation in left ventricular assist device explantation and heart transplantation.. Journal of cardiac surgery. 2022. DOI: 10.1111/jocs.16956.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.