m+Medelevator.RU
← Поиск публикаций
JACC. Cardiovascular interventions · 2025

Electrosurgical Leaflet Modification to Prevent Coronary Obstruction During Transcatheter Aortic Valve Replacement in Failing Native and Bioprosthetic Valves.

СтатьяМногоцентровое исследованиеПоддержка NIH: внешняя программаНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Rogers Toby, Khan Jaffar M, Laham Roger, Babaliaros Vasilis C, Greenbaum Adam B, Satler Lowell F, Latib Azeem, Rier Jeremy D, Harvey James E, Oldemeyer J Bradley, McCabe Jamie M, Depta Jeremiah P, Foerst Jason R, Daniels David V, Petrossian George A, Robinson Newell B, Chung William B, Byku Isida, Scotti Andrea, Strote Justin A, Schelegle Aaron R, Halaby Rim N, Bruce Christopher G, Tian Xin, Stine Annette M, Lederman Robert J, TELLTALE Investigators

Аннотация

BACKGROUND: In patients with high-risk anatomy, transcatheter aortic valve replacement (TAVR) risks coronary obstruction which is associated with high morbidity and mortality. Bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction (BASILICA) is effective at preventing coronary obstruction.

OBJECTIVES: The authors sought to test the safety and effectiveness of TELLTALE (Transmural Electrosurgery LeafLet Traversal And Laceration Equipment), the first dedicated transcatheter electrosurgery leaflet modification system.

METHODS: The TELLTALE trial (NCT05666713) enrolled participants with high or prohibitive surgical risk undergoing TAVR for native aortic stenosis or bioprosthetic valve failure and in whom computed tomography core laboratory confirmed high risk of coronary obstruction. The primary efficacy endpoint was technical success in the catheterization lab and the primary safety endpoint was inpatient safety.

RESULTS: A total of ninety participants (age 79 [75-82] years, 66% [59/90] female, 30 native aortic stenosis, 60 bioprosthetic valve failure) were enrolled at 11 centers in the United States. High risk of coronary obstruction was confirmed by computed tomography (coronary height 8.2 [6.3-10.8] mm, valve-to-coronary distance 3.4 [2.7-3.7] mm, valve-to-sinotubular-junction distance 1.2 [0.7-1.7] mm). Leaflet modification was easy and reliable with median 1 traversal and laceration attempt per participant. The primary efficacy endpoint was met in 100% (95% CI: 95%-100%) of participants. The primary safety endpoint was met in 96% (95% CI: 88%-99%) of participants, including zero mortality, 3 strokes (1 disabling, 2 nondisabling), and 1 coronary obstruction from transcatheter heart valve commissural suture pos-malalignment. At 30 days, there was zero mortality and zero delayed coronary obstruction.

CONCLUSIONS: Electrosurgical leaflet modification using the TELLTALE system is safe and effective in patients undergoing TAVR for native aortic stenosis or bioprosthetic valve failure at high risk of coronary obstruction.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты