Bilirubin Elevation Associated With Pulsed Field Ablation Induced Hemolysis in Patients With Paroxysmal Atrial Fibrillation.
Аннотация
BACKGROUND AND AIMS: Hemolysis-related renal failure after pulsed field ablation (PFA) has been described in recently published cases, we reported the incidence of bilirubin elevation after PFA utilizing a novel hexaspline PFA catheter.
METHODS
PFA was performed in patients with paroxysmal atrial fibrillation using novel hexaspline PFA catheter, and serum bilirubin, hemoglobin and renal function were measured at baseline and the next day post ablation.
RESULTS
A total of 94 patients were analyzed, and 30 of 94 (31.9%) patients had obvious total bilirubin elevation the next day post PFA. In the 30 patients, 26 (86.7%) patients had a predominantly indirect hyperbilirubinemia, suggesting a likely presence of PFA-induced hemolysis. The liver enzyme contents post ablation were normal in all patients and no signs of hemolytic anemia and renal function injury were detected. The impact factors associated with indrect hyperbilirubinemia were also analyzed and higher number of applications tented to produce PFA-induced hemolysis. More than 86.5 applications seem to have a better sensivity and specificity to predict hemolysis.
CONCLUSION
Intravascular hemolysis can occur after utilizing novel hexaspline catheter, but the severity of hemolysis was mild and temporary. The number of applications appears to be a determining factor leading to hemolysis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.