MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.athoracsur.2019.12.049

Медицинская библиотека · The Annals of thoracic surgery · 2020

HeartWare Ventricular Assist Device Cannula Position and Hemocompatibility-Related Adverse Events.

Imamura Teruhiko, Narang Nikhil, Nitta Daisuke, Fujino Takeo, Nguyen Ann, Chung Ben, Holzhauser Luise, Kim Gene, Raikhelkar Jayant, Kalantari Sara, Smith Bryan, Juricek Colleen, Rodgers Daniel, Ota Takeyoshi, Song Tae, Jeevanandam Valluvan, Sayer Gabriel, Uriel Nir

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

HeartWare ventricular assist device (HVAD) cannula position is associated with hemodynamics and heart failure readmissions. However, its impact on hemocompatibility-related adverse events (HRAEs) remains uncertain.

METHODS

HVAD patients were followed for 1 year after index hospitalization, when cannula coronal angle was quantified from chest x-ray film. Invasive right heart catheterization and transthoracic echocardiography were performed. One-year occurrences of each HRAE were compared between those with and without a cannula coronal angle of greater than 65 degrees.

RESULTS

Among 63 HVAD patients (median age 60 years, 63% male), 10 (16%) had a cannula coronal angle greater than 65 degrees. The wide-angle group had elevated intracardiac pressures and lower pulmonary artery pulsatility index (P < .05). They also had reduced right ventricular function by echocardiography. Freedom from HRAEs tended to be lower in the wide-angle group (24% vs 62%; P = .11). The rate of gastrointestinal bleeding was significantly higher in the greater than 65 degrees group (0.90 events/year vs 0.40 events/year; P = .013). The rates of stroke and pump thrombosis were statistically comparable irrespective of cannula angle (P > .05).

CONCLUSIONS

HVAD cannula coronal angle was associated with reduced right ventricular function and HRAEs. Prospective studies evaluating surgical techniques to ensure optimal device positioning and its effects on HRAEs are warranted.

Для цитирования: Imamura Teruhiko, Narang Nikhil, Nitta Daisuke, Fujino Takeo, Nguyen Ann, Chung Ben, Holzhauser Luise, Kim Gene, Raikhelkar Jayant, Kalantari Sara, Smith Bryan, Juricek Colleen, Rodgers Daniel, Ota Takeyoshi, Song Tae, Jeevanandam Valluvan, Sayer Gabriel, Uriel Nir. HeartWare Ventricular Assist Device Cannula Position and Hemocompatibility-Related Adverse Events.. The Annals of thoracic surgery. 2020. DOI: 10.1016/j.athoracsur.2019.12.049.