OBJECTIVE

We sought to determine the clinical and imaging outcomes after placement of the OptionELITE inferior vena cava (IVC) filter (Argon Medical Devices, Plano, TX).

METHODS

The clinical characteristics and imaging findings of patients who had undergone OptionELITE filter placement at six affiliated hospitals between January 1, 2013 and April 19, 2019, were analyzed. Data were obtained from the Radiology Information System, imaging database, and electronic medical records. The patients were followed up until 1 month after filter removal, death, or the first clinic visit and imaging study after June 1, 2019.

RESULTS

A total of 603 filter placement procedures in 594 patients were evaluated. Of the 603 procedures, 602 were technically successful (99.8%). Of the 594 patients, 356 had presented with acute pulmonary embolism and/or deep vein thrombosis. A total of 189 filters were retrieved from the 191 patients for whom retrieval had been attempted at 4.3 ± 6.3 months (median, 2.5 months) after placement. The 414 filters without removal procedures had been followed up for 14.8 ± 17.7 months (median, 6.9 months). During the follow-up period, 57 patients (9.5%) had developed new or worsening deep vein thrombosis caudal to the filter and 19 (3.2%) had experienced new symptomatic nonfatal pulmonary embolism after filter placement. One filter had migrated and fractured, and another filter had fractured and a strut was lost. Perforation was demonstrated equivocally in 12 (3%) and definitely in 17 (4.3%) of 399 patients with adequate imaging studies available, with penetration into surrounding structures in 13 patients (3.3%). Of 332 patients with adequate imaging studies, 12 (3.6%) had developed substantial IVC thrombosis and 4 (1.2%) had developed IVC occlusion.

CONCLUSIONS

The rates of symptomatic pulmonary embolism and filter-related complications were low after OptionELITE IVC filter placement, and filter retrieval, when attempted, was 99% successful.

Для цитирования: Albertson Nathan, Rice Mitchell, Schmitz Adam, Eskew Joseph, Haste Paul, Johnson Matthew S. Clinical and imaging outcomes of OptionELITE vena cava filter placement procedures.. Journal of vascular surgery. Venous and lymphatic disorders. 2022. DOI: 10.1016/j.jvsv.2022.08.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.