DOI 10.1016/j.ijcard.2025.133135

Медицинская библиотека · International journal of cardiology · 2025

12-month outcomes of pulsed-field ablation with a variable-diameter circular catheter for paroxysmal atrial fibrillation: Insights into conscious sedation and fluoroscopy use.

Chen Shiquan, Liu Qiang, Yin Xiaomeng, Xiao Fangyi, Wang Yuegang, Ju Weizhu, Hu Hongde, Zhang Jidong, Li Jun, Li Ruibin, Jiang Ruhong, Yu Lu, Cheng Hui, Li Jianyong, Li Xinzhong, Jiang Xiaohong, Chen Hongwu, Liu Hailei, Wang Zidun, Dai Yinyin, Li Yan, Fu Hua, Chen Minglong, Jiang Chenyang

СтатьяМногоцентровое исследованиеКлиническое исследованиеПеревод в очереди

Аннотация

BACKGROUND

Pulsed-field ablation (PFA) emerges as an innovative nonthermal energy modality for catheter ablation of atrial fibrillation (AF). This study aimed to assess the safety and effectiveness of a novel PFA system that uses a multichannel, circular ablation catheter with adjustable diameters in treating paroxysmal AF.

METHODS

This clinical trial (PF-Beat-AF) was a prospective, multicenter, single-arm study. Patients with paroxysmal AF underwent pulmonary vein isolation (PVI) using the PFA system, and were followed at discharge, 7-day, 1-, 3-, 6- and 12-month post-procedure. Primary safety endpoint was the incidence of primary adverse events (PAEs). While primary effectiveness endpoint was freedom from documented atrial arrhythmia recurrence lasting >30 s during the 3-12 months evaluation period.

RESULTS

A total of 161 patients were enrolled and 159 patients were treated in eight centers. Conscious sedation was used in 68.6 % of cases, and 40.9 % of procedures were completed without fluoroscopy. The mean total procedure time was 132.3 ± 40.4 min, and fluoroscopy time averaged 6.4 ± 6.8 min. Acute PVI was achieved in 100 % of patients. One PAE (0.6 %) occurred (pericardial tamponade, resolved). At 12 months, 87.7 % (95 % CI: 82.5 %-92.9 %) of patients remained free from atrial arrhythmia after the blanking period. No significant differences in effectiveness were observed between conscious sedation and general anesthesia/deep sedation, or fluoroscopy and zero-fluoroscopy cases.

CONCLUSION

The results of clinical trial demonstrated the safety and effectiveness of the PFA system in treating paroxysmal AF, including successful use with conscious sedation and zero-fluoroscopy.

Для цитирования: Chen Shiquan, Liu Qiang, Yin Xiaomeng, Xiao Fangyi, Wang Yuegang, Ju Weizhu, Hu Hongde, Zhang Jidong, Li Jun, Li Ruibin, Jiang Ruhong, Yu Lu, Cheng Hui, Li Jianyong, Li Xinzhong, Jiang Xiaohong, Chen Hongwu, Liu Hailei, Wang Zidun, Dai Yinyin, Li Yan, Fu Hua, Chen Minglong, Jiang Chenyang. 12-month outcomes of pulsed-field ablation with a variable-diameter circular catheter for paroxysmal atrial fibrillation: Insights into conscious sedation and fluoroscopy use.. International journal of cardiology. 2025. DOI: 10.1016/j.ijcard.2025.133135.