DOI 10.5435/jaaos-d-23-00590

Медицинская библиотека · The Journal of the American Academy of Orthopaedic Surgeons · 2023

The Biomechanical Effects of Intramedullary Fixation to the Level of the Physeal Scar on Distal Tibia Fracture Stability.

Schneiderman Brian A, Oppizzi Giovanni, Gupta Jayesh, Burke Cynthia E, Munn Murty, Wilkinson Brandon, Zhang Li-Qun, Jaeblon Todd, O'Toole Robert V, Sciadini Marcus F

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

The purpose of this study was to evaluate whether intramedullary nail contact with physeal scar improves construct mechanics when treating distal tibial shaft fractures.

METHODS

Axially unstable extra-articular distal tibia fractures were created in 30 fresh frozen cadaveric specimens (15 pairs, mean age 79 years). Specimens underwent intramedullary nailing to the level of the physeal scar locked with one or two interlocks or short of the physeal scar locked with two interlocks (reference group). Specimens were subjected to 800N of axial load for 25,000 cycles. Primary outcomes were stiffness before and after cyclic loading. Secondary outcomes were load to failure, load at 3 mm displacement, plastic deformation, and total deformation.

RESULTS

The physeal scar with one interlock cohort demonstrated 3.8% greater stiffness before cycling ( P = 0.75) and 1.7% greater stiffness after cycling ( P = 0.86) compared with the reference group. The physeal scar with two interlocks group exhibited 0.3% greater stiffness before cycling ( P = 0.98) and 8.4% greater stiffness after cycling ( P = 0.41) in relation to the reference group. No differences were identified regarding load to failure or load at 3 mm displacement. In specimens with two interlocks, those in contact with the physeal scar demonstrated significantly less plastic ( P = 0.02) and total ( P = 0.04) deformation.

CONCLUSIONS

Constructs ending at the physeal scar demonstrated stiffness and load to failure similar to those without physeal scar contact. Less plastic and total deformation was noted in two-interlock constructs with physeal scar contact, suggesting a possible protective effect provided by the physeal scar. These data argue that physeal scar contact may offer a small mechanical benefit in nailing distal tibia fractures, but clinical relevance remains unknown.

Для цитирования: Schneiderman Brian A, Oppizzi Giovanni, Gupta Jayesh, Burke Cynthia E, Munn Murty, Wilkinson Brandon, Zhang Li-Qun, Jaeblon Todd, O'Toole Robert V, Sciadini Marcus F. The Biomechanical Effects of Intramedullary Fixation to the Level of the Physeal Scar on Distal Tibia Fracture Stability.. The Journal of the American Academy of Orthopaedic Surgeons. 2023. DOI: 10.5435/jaaos-d-23-00590.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.