Impact of heart rate, aortic compliance and stroke volume on the aortic regurgitation fraction studied in an ex vivo pig model.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Reil Jan Christian, Saisho Hiroyuki, Jockwer Antonia, Fujita Buntaro, Paluszkiewicz Lech, Reil Gert-Hinrich, Ensminger Stephan, Scharfschwerdt Michael, Aboud Anas

Аннотация

INTRODUCTION: Drug therapy to reduce the regurgitation fraction (RF) of high-grade aortic regurgitation (AR) by increasing heart rate (HR) is generally recommended. However, chronic HR reduction in HFREF patients can significantly improve aortic compliance and thereby potentially decrease RF. To clarify these contrasts, we examined the influence of HR, aortic compliance and stroke volume (SV) on RF in an ex vivo porcine model of severe AR.

METHODS: Experiments were performed on porcine ascending aorta with aortic valves (n=12). Compliance was varied by inserting a Dacron graft close to the aortic valve. Both tube systems were connected to a left heart simulator varying HR and SV. AR was accomplished by punching a 0.3 cm2 hole in one aortic cusp. Flow, RF, SV and aortic pressure were measured, aortic compliance with transoesophageal ultrasound probes.

RESULTS: Compliance of the aorta was significantly reduced after Dacron graft insertion (0.55%±0.21%/mm Hg vs 0.01%±0.007%/mm Hg, p<0.001, respectively). With increasing HR, RF was significantly reduced in each steady state of the native aorta (HR 40 bpm: 88%±7% vs HR 120 bpm: 42%±10%; p<0.001), but Dacron tube did not affect RF (HR 40 bpm: 87%±8%; p=0.79; HR 120 bpm: 42%±3%; p=0.86). Increasing SV also reduced RF independent of the stiff Dacron graft.

CONCLUSION: Aortic compliance did not affect AR in the ex vivo porcine model of AR. RF was significantly reduced with increasing HR and SV. These results affirm that HR lowering and negative inotropic drugs should be avoided to treat severe AR.