European heart journal. Case reports · 2024
Hybrid convergent procedure with proactive oesophageal cooling for the treatment of long-standing persistent atrial fibrillation: a case series.
Клинический случайСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Velasco Alejandro, Buch Chirag, Hui Dawn, Joseph Christopher, Onsager David, Zagrodzky William, Kulstad Erik, Nayak Hemal M
Аннотация
BACKGROUND: The hybrid convergent procedure is approved to treat symptomatic patients with long-standing persistent atrial fibrillation (AF). Despite direct visualization during surgical ablation as well as the use of luminal oesophageal temperature (LET) monitoring, oesophageal injury is still possible. A dedicated device for proactive oesophageal cooling has recently been cleared by the Food and Drug Administration to reduce the likelihood of ablation-related oesophageal injury resulting from radiofrequency cardiac ablation procedures. This report describes the first uses of proactive oesophageal cooling for oesophageal protection during the epicardial ablation portion of hybrid convergent procedures.
CASE SUMMARY: Five patients with long-standing persistent AF underwent hybrid convergent ablations with the use of proactive oesophageal cooling as means of oesophageal protection. All cases were completed successfully with no adverse effects. Most notably, cases were shorter when compared to cases using LET monitoring, likely due to lack of pauses for overheating of the oesophagus that would otherwise be required to prevent damage to the oesophagus.
DISCUSSION: This report describes the first uses of proactive oesophageal cooling for oesophageal protection during the epicardial ablation portion of five hybrid convergent procedures. Use of cooling enabled uninhibited deployment of lesions without the need to pause energy delivery due to elevated temperatures in the oesophagus, providing a feasible alternative to LET monitoring.
CASE SUMMARY: Five patients with long-standing persistent AF underwent hybrid convergent ablations with the use of proactive oesophageal cooling as means of oesophageal protection. All cases were completed successfully with no adverse effects. Most notably, cases were shorter when compared to cases using LET monitoring, likely due to lack of pauses for overheating of the oesophagus that would otherwise be required to prevent damage to the oesophagus.
DISCUSSION: This report describes the first uses of proactive oesophageal cooling for oesophageal protection during the epicardial ablation portion of five hybrid convergent procedures. Use of cooling enabled uninhibited deployment of lesions without the need to pause energy delivery due to elevated temperatures in the oesophagus, providing a feasible alternative to LET monitoring.