MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2147/mder.s328485

Медицинская библиотека · Medical devices (Auckland, N.Z.) · 2021

Invention and Pilot Study of the Efficacy and Safety of the SUPRAtube Device in Continuous Supraglottic Aspiration for Intubated and Mechanically Ventilated Patients.

Ramírez-Sarmiento Alba, Aya Orlando, Cáceres-Rivera Diana, Reyes Carlos F, Espitia Angela, Pizarro Camilo, Gea Joaquim, Castillo Victor R, Orozco-Levi Mauricio

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Bronchoaspiration of content that accumulates in the supraglottic area (eg, saliva, gastroesophageal reflux) is a risk factor for ventilator-associated pneumonia. A continuous supraglottic suction system may decrease the risk of bronchoaspiration in these patients.

OBJECTIVE

(1) Constructing a conceptual model and functional prototype of a continuous supraglottic suction device for use in humans; (2) defining functional characteristics in ex vivo swine head models; and (3) evaluating its efficacy and safety in mechanically ventilated patients.

METHODS

Study conducted in three phases. First phase: definition of distances and diameters of the triangle determined by dental arch, posterior oropharynx and vallecula, and diameter of the oropharynx in axial projection; and identification of the declining area of supraglottic suction. Second phase: design engineering and functional prototype evaluated in ex vivo models. Third phase: evaluation of device use in terms of safety and efficacy in ventilated patients.

RESULTS

We obtained a final functional model of the SUPRAtube device injected into PVC for medical use. Device effectiveness in in vitro simulation showed a high and fast suction capacity of liquid and thick volumes. Study of swine heads allowed to validate the shape, size and functional fenestration of the device. Study in intubated and mechanically ventilated patients showed a high supraglottic suction capacity and the absence of local adverse events during 72 (7-240) hours of continuous operation.

CONCLUSION

Our study describes the process of conceptualization, design and production of a practical, safe, low-cost continuous supraglottic suction device without representing antibiotic pressure, which appears to be a new complementary preventive strategy for the standard management of intubated and mechanically ventilated patients.

Для цитирования: Ramírez-Sarmiento Alba, Aya Orlando, Cáceres-Rivera Diana, Reyes Carlos F, Espitia Angela, Pizarro Camilo, Gea Joaquim, Castillo Victor R, Orozco-Levi Mauricio. Invention and Pilot Study of the Efficacy and Safety of the SUPRAtube Device in Continuous Supraglottic Aspiration for Intubated and Mechanically Ventilated Patients.. Medical devices (Auckland, N.Z.). 2021. DOI: 10.2147/mder.s328485.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.