Hemodynamic Characterization of Peripheral Arterio-Venous Malformations Using Rapid Contrast-Enhanced MR Imaging: An In Vitro and In Vivo Study.

Calastra Camilla Giulia, Bono Marika, Granada Aloma Blanch, Tuleja Aleksandra, Bernhard Sarah Maike, Diaz-Zuccarini Vanessa, Balabani Stavroula, Obrist Dominik, von Tengg-Kobligk Hendrik, Jung Bernd

СтатьяПеревод в очереди

Аннотация

PURPOSE: Peripheral arterio-venous malformations (pAVMs) are vascular defects often requiring extensive medical treatment. To improve disease management, hemodynamic markers based on 2D Digital Subtraction Angiography (DSA) data were previously defined to classify pAVMs. However, DSA offers only 2D information, involves ionizing radiation, and requires intra-arterial intervention. We hypothesized that pAVMs could be classified with the same approach with 3D dynamic contrast-enhanced MR-based data. To this end, the present work aims to develop a computational classification system for pAVMs using 3D dynamic contrast-enhanced MR-based data.

METHODS

A pAVM phantom was imaged using both DSA and MRI to validate the methodology, which was then applied to 10 MR-based in vivo datasets. A semi-automated vessel detection algorithm, based on the standard deviation of each voxel or pixel in time, was used. Classification was performed by identifying the time of arrival (CAToA) of contrast agent (CA) and the maximum time derivative of the CA transport in each pixel or voxel (CAsi).

RESULTS

Normalized CAToA and CAsi histograms showed no significant difference between in vitro DSA and MRI (respectively χ2 = 0.20, p = 0.65 and χ2 = 0.21, p = 0.65), validating the methodology to classify pAVMs. CAToA histograms for type II-IV AVMs derived from in vivo MR-based data aligned with DSA patterns and known hemodynamics. CAToA histograms of capillary-venulous AVMs were distinct, with non-zero values at later times than other AVM types, representing late venous drainage. Type IV AVMs histograms for CAsi were more right-skewed than those derived from types II and III pAVMs.

CONCLUSIONS

MR image quality and temporal resolution are sufficient to allow a classification of pAVMs. This classification method has the potential to become a diagnostic tool for the surgical navigation of pAVMs for clinicians.

Для цитирования: Calastra Camilla Giulia, Bono Marika, Granada Aloma Blanch, Tuleja Aleksandra, Bernhard Sarah Maike, Diaz-Zuccarini Vanessa, Balabani Stavroula, Obrist Dominik, von Tengg-Kobligk Hendrik, Jung Bernd. Hemodynamic Characterization of Peripheral Arterio-Venous Malformations Using Rapid Contrast-Enhanced MR Imaging: An In Vitro and In Vivo Study.. Annals of biomedical engineering. 2025. DOI: 10.1007/s10439-025-03766-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.