First-in-human trial of atrial fibrillation ablation using real-time tissue optical assessment to predict pulsed field lesion durability.
Аннотация
AIMS: Loss of bipolar electrograms immediately after pulsed field ablation (PFA) makes lesion durability assessment challenging.
OBJECTIVE
The aim of this trial (NCT06700226) was to evaluate a novel ablation system that can optically predict lesion durability by detecting structural changes in the tissue during ablation.
METHODS AND
RESULTS
Patients with paroxysmal atrial fibrillation underwent pulmonary vein isolation (PVI) using PFA (AblaView®, MedLumics). Using polarization-sensitive optical coherence reflectometry (PS-OCR), reflective characteristics of myocardial tissue and visualization of real-time contrast between healthy tissue and ablated tissue using a drop in tissue birefringence (BiR) was assessed. Wide antral PVI was performed using single point irrigated PFA (unipolar, 1800V, 3 trains, 21 s). Remapping was performed at 3 months. Primary efficacy outcome was the ability of PS-OCR to predict lesion durability at 3-month remapping. Serious adverse events were recorded. Ten patients were included. In total, 38/40 PVs could be isolated with the system. The mean drop of BiR was 17.3 ± 11.5%. Dragging across the ablation lines showed a persistent drop in BiR. During the remap procedures (8/10 patients ablated only with PFA), 12 PVs (37.5%) were found to be electrically reconnected. The mean loss of BiR during all PFA for durable lesions was 20.9%, while only 10.1% BiR loss was observed during the index ablation for reconnected areas (P < 0.001). None of the points with ≥17% loss of birefringence was found to be reconnected.
CONCLUSION
This first-in-human study supports the use of real-time drop in tissue BiR for lesion assessment and durability during PFA delivery, and its procedural safety.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.