Clinical Study of High-Resolution C-Arm CT in Mechanical Recanalization and Stent Implantation for Chronic Cerebral Artery Occlusion.
Аннотация
BACKGROUND
In recent years, among patients with chronic cerebral artery occlusion, recanalization can be achieved by an endovascular operation. However, complications and restenosis rates remain high.
OBJECTIVE
To evaluate the utility of high-resolution C-arm CT (Dyna micro-CT) for stent placement in patients with chronic cerebral artery occlusion.
METHODS AND MATERIALS: We retrospectively reviewed the clinical data of 27 patients with chronic cerebral artery occlusion who underwent mechanical recanalization and stent implantation. Images were reconstructed using conventional C-arm CT (Dyna CT) and Dyna micro-CT. Whether the stent was fully expanded and image quality was evaluated. Follow-up assessments included clinical and angiographic outcomes and complications.
RESULTS
Twenty-two patients successfully underwent stenting (22 stents; 14 cases: Neuroform EZ; eight cases: Enterprise); stenting failed in five patients. Compared to Dyna CT, Dyna micro-CT afforded improved visualization of the stent structure, providing significantly improved image quality (P < 0.05). In seven patients, the stent under-expanded and dilatation was performed; thereafter, stent malapposition improved. One patient experienced sudden headache 22 hours after the procedure; CT showed intraparenchymal hemorrhage. The remaining 21 patients did not have acute thrombosis or bleeding complications and were followed up by imaging for 3-6 months. In three patients, digital subtraction angiography showed mild in-stent stenosis.
CONCLUSIONS
High-resolution C-arm CT can improve visualization of stent structures in chronic cerebral artery occlusion, making it easy to determine the extent of stent deployment and potentially reduce complications and stent restenosis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.