MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.apmr.2024.08.026

Медицинская библиотека · Archives of physical medicine and rehabilitation · 2024

Evaluating the Effectiveness of Transtibial Prosthetic Socket Shape Design Using Artificial Intelligence: A Clinical Comparison With Traditional Plaster Cast Socket Designs.

van der Stelt Merel, Berends Bo, Papenburg Marco, Langenhuyzen Tom, Maal Thomas, Brouwers Lars, de Jong Guido, Leijendekkers Ruud

СтатьяСравнительное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To investigate the feasibility of creating an artificial intelligence (AI) algorithm to enhance prosthetic socket shapes for transtibial prostheses, aiming for a less operator-dependent, standardized approach.

DESIGN

The study comprised 2 phases: first, developing an AI algorithm in a cross-sectional study to predict prosthetic socket shapes. Second, testing the AI-predicted digitally measured and standardized designed (DMSD) prosthetic socket against a manually measured and designed (MMD) prosthetic socket in a 2-week within-subject cross-sectional study.

SETTING

The study was done at the rehabilitation department of the Radboud University Medical Center in Nijmegen, the Netherlands.

PARTICIPANTS

The AI algorithm was developed using retrospective data from 116 patients from a Dutch orthopedic company, OIM Orthopedie, and tested on 10 randomly selected participants from Papenburg Orthopedie.

INTERVENTIONS: Utilization of an AI algorithm to enhance the shape of a transtibial prosthetic socket.

MAIN OUTCOME MEASURES: The algorithm was optimized to minimize the error in the test set. Participants' socket comfort score and fitting ratings from an independent physiotherapist and prosthetist were collected.

RESULTS

Predicted prosthetic shapes deviated by 2.51 mm from the actual designs. In total, 8 of 10 DMSD and all 10 MMD-prosthetic sockets were satisfactory for home testing. Participants rated DMSD-prosthetic sockets at 7.1 ± 2.2 (n=8) and MMD-prosthetic sockets at 6.6 ± 1.2 (n=10) on average.

CONCLUSIONS

The study demonstrates promising results for using an AI algorithm in prosthetic socket design, but long-term effectiveness and refinement for improved comfort and fit in more deviant cases are necessary.

Для цитирования: van der Stelt Merel, Berends Bo, Papenburg Marco, Langenhuyzen Tom, Maal Thomas, Brouwers Lars, de Jong Guido, Leijendekkers Ruud. Evaluating the Effectiveness of Transtibial Prosthetic Socket Shape Design Using Artificial Intelligence: A Clinical Comparison With Traditional Plaster Cast Socket Designs.. Archives of physical medicine and rehabilitation. 2024. DOI: 10.1016/j.apmr.2024.08.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.