DOI 10.18502/ijm.v14i6.11254

Медицинская библиотека · Iranian journal of microbiology · 2022

Design of an optical nanobiosensor for detection of Legionella pneumophila in water samples.

Karimiravesh Raheleh, Mohabati Mobarez Ashraf, Behmanesh Mehrdad, Nikkhah Maryam, Talebi Bezmin Abadi Amin, Esmaeilli Saber

СтатьяПеревод в очереди

Аннотация

BACKGROUND AND

OBJECTIVES

Legionella spp. is a causative agent of Legionnaires' disease that creates public health problems. Isolation of these bacteria from water sources is essential to identify outbreak origins and prevent disease. Diagnostic biosensors for water quality control to protect consumers from water-borne infections can predict many outbreaks. Gold nanoparticles conjugated probes are a new generation of diagnostic tools. In this study, an optical nano biosensor was designed and characterized to detect Legionella pneumophila in water samples rapidly.

MATERIALS AND

METHODS

Thiolated probes designed for the mip gene were attached to gold nanoparticles and then water samples containing Legionella pneumophila were examined.

RESULTS

The limit of detection for PCR and biosensor was 104 and 103 copy numbers/μl, respectively. Biosensor sensitivity and PCR were reported to be 90% (18 out of 20) and 85% (17 out of 20), respectively. Specificity 100% has been reported for both methods.

CONCLUSION

According to the obtained results, this method has the potential to diagnose L. pneumophila with high sensitivity and specificity. This system can be employed as a practical tool for rapid, accurate, high-sensitivity, and acceptable detection of Legionella pneumophila in contaminated water, which is cost-effective in terms of cost and time.

Для цитирования: Karimiravesh Raheleh, Mohabati Mobarez Ashraf, Behmanesh Mehrdad, Nikkhah Maryam, Talebi Bezmin Abadi Amin, Esmaeilli Saber. Design of an optical nanobiosensor for detection of Legionella pneumophila in water samples.. Iranian journal of microbiology. 2022. DOI: 10.18502/ijm.v14i6.11254.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.