DOI 10.1016/j.jse.2014.10.013

Медицинская библиотека · Journal of shoulder and elbow surgery · 2014

Patient-specific targeting guides compared with traditional instrumentation for glenoid component placement in shoulder arthroplasty: a multi-surgeon study in 70 arthritic cadaver specimens.

Throckmorton Thomas W, Gulotta Lawrence V, Bonnarens Frank O, Wright Stephen A, Hartzell Jeffrey L, Rozzi William B, Hurst Jason M, Frostick Simon P, Sperling John W

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

HYPOTHESIS AND

BACKGROUND

The purpose of this study was to compare the accuracy of patient-specific guides for total shoulder arthroplasty (TSA) with traditional instrumentation in arthritic cadaver shoulders. We hypothesized that the patient-specific guides would place components more accurately than standard instrumentation.

MATERIALS AND

METHODS

Seventy cadaver shoulders with radiographically confirmed arthritis were randomized in equal groups to 5 surgeons of varying experience levels who were not involved in development of the patient-specific guidance system. Specimens were then randomized to patient-specific guides based off of computed tomography scanning, standard instrumentation, and anatomic TSA or reverse TSA. Variances in version or inclination of more than 10° and more than 4 mm in starting point were considered indications of significant component malposition.

RESULTS

TSA glenoid components placed with patient-specific guides averaged 5° of deviation from the intended position in version and 3° in inclination; those with standard instrumentation averaged 8° of deviation in version and 7° in inclination. These differences were significant for version (P = .04) and inclination (P = .01). Multivariate analysis of variance to compare the overall accuracy for the entire cohort (TSA and reverse TSA) revealed patient-specific guides to be significantly more accurate (P = .01) for the combined vectors of version and inclination. Patient-specific guides also had fewer instances of significant component malposition than standard instrumentation did.

CONCLUSION

Patient-specific targeting guides were more accurate than traditional instrumentation and had fewer instances of component malposition for glenoid component placement in this multi-surgeon cadaver study of arthritic shoulders. Long-term clinical studies are needed to determine if these improvements produce improved functional outcomes.

Для цитирования: Throckmorton Thomas W, Gulotta Lawrence V, Bonnarens Frank O, Wright Stephen A, Hartzell Jeffrey L, Rozzi William B, Hurst Jason M, Frostick Simon P, Sperling John W. Patient-specific targeting guides compared with traditional instrumentation for glenoid component placement in shoulder arthroplasty: a multi-surgeon study in 70 arthritic cadaver specimens.. Journal of shoulder and elbow surgery. 2014. DOI: 10.1016/j.jse.2014.10.013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.