Reduction of scapulohumeral subluxation with posterior augmented glenoid implants in anatomic total shoulder arthroplasty: Short-term 3D comparison between pre- and post-operative CT.

Terrier Alexandre, Goetti Patrick, Becce Fabio, Farron Alain

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Failure rates in anatomic total shoulder arthroplasty (aTSA) are higher in case of asymmetric glenoid bone loss secondary to posterior wear, and in persistent static posterior subluxation of the humeral head (PSH).

HYPOTHESIS: This study aimed to test the hypothesis that the combined use of posterior augmented glenoid (PAG) implants with three-dimensional (3D) surgical planning and patient-specific instrumentation (PSI) guides helps reduce short-term PSH after aTSA in patients with type B2-B3 glenoids.

PATIENTS AND

METHODS

We included nine consecutive patients with primary glenohumeral osteoarthritis and type B2 or B3 glenoids, who underwent aTSA with cemented keeled PAG implants (posterior augments of 15, 25, or 35 degrees). All patients underwent preoperative shoulder computed tomography (CT) scans, with 3D surgical planning coupled to PSI at the time of surgery. Postoperative shoulder CT scans were performed at an average of 14 weeks (range, 10-21 weeks). Scapulohumeral subluxation and glenoid version and inclination were measured in 3D, on both pre- and post-operative CT scans, using the same reliable quantitative method.

RESULTS

There was a significant decrease in scapulohumeral subluxation from 49±12% preoperatively to 22±17% postoperatively (p=0.0039), with a large effect size (Cohen's d=1.89). Preoperative glenoid version was corrected from -17.3±9.4 degrees to -5.2±7.5 degrees postoperatively. The absolute difference between the postoperative and surgically planned version and inclination was 5.4±3.6 degrees and 3.3±2.0 degrees, respectively.

DISCUSSION: The combined use of PAG implants with 3D preoperative planning and PSI results in a significant decrease in short-term PSH and glenoid version in patients with asymmetric posterior glenoid wear. We suggest that such implants should not be limited to posterior augmentation, because glenoid deficiency can also be observed in other glenoid sectors.

LEVEL OF EVIDENCE: IV, Basic science study.

Для цитирования: Terrier Alexandre, Goetti Patrick, Becce Fabio, Farron Alain. Reduction of scapulohumeral subluxation with posterior augmented glenoid implants in anatomic total shoulder arthroplasty: Short-term 3D comparison between pre- and post-operative CT.. Orthopaedics & traumatology, surgery & research : OTSR. 2020. DOI: 10.1016/j.otsr.2020.03.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.