Load transfer after cemented total shoulder arthroplasty.
Аннотация
BACKGROUND
Glenoid loosening is the primary reason for failure after a total shoulder arthroplasty (TSA), but the failure mechanism is not yet known. This study determined how the load transfer and stress distribution are affected by the introduction of a glenoid implant.
METHODS
We developed a finite-element model of a scapula with and without a virtually implanted modern glenoid prosthesis design. Two load magnitudes were considered: normal and high. Loading locations were simulated at the center and at 4 eccentric positions on the glenoid. A metal-backed implant was also simulated to understand the effect of fixation stiffness.
RESULTS
In the intact glenoid, for both center and eccentric loading, the majority of stress was distributed in the cancellous bone, whereas after a reconstruction, stresses in that region were lower. Metal-backed implants further decreased the joint load carried by the bone. Stresses in the cement layer increased during eccentric and high-magnitude loading.
CONCLUSION
This study provided a basic understanding of the load-sharing phenomenon after a TSA that could explain glenoid loosening failure. Our results suggest that with reconstruction of the glenoid with a contemporary implant, the load transfer pattern is significantly altered, with eccentric and high-magnitude loads increasing stresses in the cement indicating potential for failure. The use of a metal-backed implant reduces the load carried by the bone, which may be detrimental to long-term TSA survival.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.