Breast radiation dose with contrast-enhanced mammography-guided biopsy: a retrospective comparison with stereotactic and tomosynthesis guidance.

Alcantara Rodrigo, Azcona Javier, Pitarch Mireia, Arenas Natalia, Castells Xavier, Milioni Pablo, Iotti Valentina, Besutti Giulia

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

This retrospective study aimed to compare the average glandular dose (AGD) per acquisition in breast biopsies guided by contrast-enhanced mammography (CEM), conventional stereotactic breast biopsy (SBB), and digital breast tomosynthesis (DBT). The study also investigated the influence of compressed breast thickness (CBT) and density on AGD. Furthermore, the study aimed to estimate the AGD per procedure for each guidance modality.

METHODS

The study included 163 female patients (mean age 57 ± 10 years) who underwent mammography-guided biopsies using SBB (9%), DBT (65%), or CEM (26%) guidance. AGD and CBT data were extracted from DICOM headers, and breast density was visually assessed. Statistical analyses included two-sample t-tests and descriptive statistics.

RESULTS

Mean AGD per acquisition varied slightly among CEM (1.48 ± 0.22 mGy), SBB (1.49 ± 0.40 mGy), and DBT (1.55 ± 0.47 mGy), with CEM presenting higher AGD at lower CBTs and less dose escalation at higher CBTs. For CBT > 55 mm, CEM showed reduced AGD compared to SBB and DBT (p < 0.001). Breast density had minimal impact on AGD, except for category A. The estimated AGD per procedure was approximately 11.84 mGy for CEM, 11.92 mGy for SBB, and 6.2 mGy for DBT.

CONCLUSION

The study found mean AGD per acquisition to be similar for CEM and SBB, with DBT slightly higher. CEM demonstrated higher AGD at lower CBT but lower AGD at higher CBT, indicating reduced dose escalation with increasing thickness. While breast density had minimal overall impact, variations were noted in category A. DBT was more dose-efficient per procedure due to fewer acquisitions required.

CLINICAL RELEVANCE STATEMENT: CEM guidance provides effective lesion visualization within safe radiation limits, improving the precision of percutaneous image-guided breast interventions and supporting its potential consideration in a wider range of breast diagnostic procedures.

KEY POINTS: Limited data exist on the AGD using CEM guidance for breast biopsies. CEM and SBB exhibit similar AGD per acquisition; DBT demonstrated the lowest AGD per procedure. Radiation from CEM guidance fits within safe limits for percutaneous image-guided breast interventions.

Для цитирования: Alcantara Rodrigo, Azcona Javier, Pitarch Mireia, Arenas Natalia, Castells Xavier, Milioni Pablo, Iotti Valentina, Besutti Giulia. Breast radiation dose with contrast-enhanced mammography-guided biopsy: a retrospective comparison with stereotactic and tomosynthesis guidance.. European radiology. 2024. DOI: 10.1007/s00330-024-10920-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.