MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.radi.2026.103331

Медицинская библиотека · Radiography (London, England : 1995) · 2026

Critical organ dose evaluation during interventional stroke imaging: Implications for radiation safety.

Özgür E, Tunçman Kayaokay D, Özgür D S, Tünay A, Günay O, Kesmezacar F F, Demir M, ALMisned G, Tekin H O

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Neurointerventional stroke imaging involves prolonged fluoroscopy and multiple angiographic acquisitions, potentially delivering meaningful radiation doses to radiosensitive head and neck organs. This study quantified organ-specific absorbed doses under clinically realistic conditions.

METHODS

Organ doses were measured using calibrated MTS-100 LiF:Mg,Ti thermoluminescent dosimeters (TLD-100) placed in an Alderson Rando anthropomorphic phantom. Simulated stroke imaging was performed on a Siemens Artis Zee C-arm system using routine clinical parameters. TLDs were calibrated under RQR-9 beam quality at an accredited SSDL and read with a Harshaw 4500 reader.

RESULTS

The highest absorbed doses occurred in the left parotid gland and C1-C2 vertebral level. Bilateral thyroid doses averaged ∼4.6 mGy, and eye lens doses ranged from 5.6 to 11.5 mGy. Dose distribution showed marked spatial variability, reflecting projection geometry and scatter contributions.

CONCLUSION

Although measured single-session doses were below deterministic thresholds, cumulative exposure during repeated or prolonged interventions may become clinically significant.

IMPLICATIONS FOR PRACTICE: The findings support the need for strict collimation, optimized projection angles, pulsed fluoroscopy, and continuous dose monitoring to minimize patient and staff exposure while maintaining diagnostic image quality.

Для цитирования: Özgür E, Tunçman Kayaokay D, Özgür D S, Tünay A, Günay O, Kesmezacar F F, Demir M, ALMisned G, Tekin H O. Critical organ dose evaluation during interventional stroke imaging: Implications for radiation safety.. Radiography (London, England : 1995). 2026. DOI: 10.1016/j.radi.2026.103331.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.