Evaluating the Efficacy of Thin Convex-probe Endobronchial Ultrasound Bronchoscope in Cadaveric Models.
Аннотация
BACKGROUND
The convex-probe endobronchial ultrasound (CP-EBUS) bronchoscope is widely used in clinical practice. Despite improvements, the existing CP-EBUS remains limited in accessing areas beyond the mediastinum due to its large distal end diameter, long rigid tip segment, and oblique viewing angle. To address this limitation, Olympus Medical Systems Corporation developed the thin CP-EBUS (TCP-EBUS). This study aimed to compare the accessibility and puncturing ability of TCP-EBUS with existing CP-EBUS in cadaveric models.
METHODS
Ten bronchoscopists conducted this study using 2 cadaveric models. The accessibility of TCP-EBUS was evaluated based on the number of bronchial generations that could be reached. To assess the puncturing ability of TCP-EBUS, needle punctures were performed on previously created simulated lesions in the segmental or subsegmental bronchial area, and puncture success rates were calculated.
RESULTS
TCP-EBUS demonstrated greater accessibility than CP-EBUS in all segmental bronchi, with statistically significant differences observed in several bronchi (P<0.05). Puncture success rates for simulated lesions using TCP-EBUS were also significantly higher than those using CP-EBUS in both the segmental (85.0% vs. 60.0%, P<0.001) and subsegmental bronchial areas (84.4% vs. 38.9%, P<0.001). In a questionnaire survey, TCP-EBUS was perceived as significantly superior in terms of usability (P<0.05).
CONCLUSION
TCP-EBUS has significantly improved both accessibility and puncture performance, providing an advantage over CP-EBUS in segmental and subsegmental bronchial areas. TCP-EBUS has the potential to expand the indications for endobronchial ultrasound-guided transbronchial needle aspiration.