MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jclinane.2012.04.007

Медицинская библиотека · Journal of clinical anesthesia · 2012

Comparison of the Pentax-AWS airway scope with the Macintosh laryngoscope for nasotracheal intubation: a randomized, prospective study.

Suzuki Akihiro, Onodera Yoshiko, Mitamura Sayuri M, Mamiya Keiko, Kunisawa Takayuki, Takahata Osamu, Henderson John J, Iwasaki Hiroshi

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

OBJECTIVE

To evaluate the effectiveness of the Pentax-AWS Airway Scope (AWS) in comparison to the Macintosh laryngoscope during nasotracheal intubation.

DESIGN

Prospective randomized study.

SETTING

Operating room of a university-affiliated hospital.

PATIENTS: 90 ASA physical status 1 and 2 adults, aged 18 to 72 years, scheduled for orthodontia surgery requiring nasotracheal intubation.

INTERVENTIONS: Patients were randomly assigned to three groups to undergo tracheal intubation with a Macintosh laryngoscope (Group Mac; n = 30), AWS with its tip inserted into the vallecula for indirect elevation of the epiglottis (Group AWS-I; n = 30), or AWS with its tip positioned posterior to the epiglottis for direct elevation of the epiglottis (Group AWS-D; n = 30).

MEASUREMENTS: Percentage of glottic opening (POGO) score at the time of laryngeal exposure, time required for intubation, and intubation difficulty scale (IDS) were measured. The frequency of postoperative sore throat and hoarseness also were noted.

MAIN

RESULTS

Patient demographics did not differ among the groups. In Groups AWS-I and AWS-D, IDS scores were reduced significantly, and the percentages of glottic opening were significantly improved, compared with the Macintosh group. Time to place the endotracheal tube was significantly shortest in Group AWS-I. In one case from each group, intubation within two attempts failed and a different approach was required.

CONCLUSION

The AWS offers better intubation conditions than the Macintosh laryngoscope during nasotracheal intubation. The AWS may be used to elevate the epiglottis both directly and indirectly for nasotracheal intubation.

Для цитирования: Suzuki Akihiro, Onodera Yoshiko, Mitamura Sayuri M, Mamiya Keiko, Kunisawa Takayuki, Takahata Osamu, Henderson John J, Iwasaki Hiroshi. Comparison of the Pentax-AWS airway scope with the Macintosh laryngoscope for nasotracheal intubation: a randomized, prospective study.. Journal of clinical anesthesia. 2012. DOI: 10.1016/j.jclinane.2012.04.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.