Robotic system for top to bottom MRgFUS therapy of multiple cancer types.
Аннотация
BACKGROUND
A robotic system for Magnetic Resonance guided Focussed Ultrasound (MRgFUS) therapy of tumours in the breast, bone, thyroid, and abdomen was developed.
METHODS
A special C-shaped structure was designed to be attached to the table of conventional magnetic resonance imaging (MRI) systems carrying 4 computer-controlled motion stages dedicated to positioning a 2.75 MHz spherically focussed transducer relative to a patient placed in the supine position. The developed system was evaluated for its MRI compatibility and heating abilities in agar-based phantoms and freshly excised tissue.
RESULTS
Compatibility of the system with a clinical high-field MRI scanner was demonstrated. FUS heating in the phantom was successfully monitored by magnetic resonance thermometry without any evidence of magnetically induced phenomena. Cigar-shaped discrete lesions and well-defined areas of overlapping lesions were inflicted in excised tissue by robotic movement along grid patterns.
CONCLUSIONS
The developed MRgFUS robotic system was proven safe and efficient by ex-vivo feasibility studies.