MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/bmj-2023-078581

Медицинская библиотека · BMJ (Clinical research ed.) · 2024

Diagnostic accuracy of magnetically guided capsule endoscopy with a detachable string for detecting oesophagogastric varices in adults with cirrhosis: prospective multicentre study.

Jiang Xi, Pan Jun, Xu Qing, Song Yu-Hu, Sun Huan-Huan, Peng Cheng, Qi Xiao-Long, Qian Yang-Yang, Zou Wen-Bin, Yang Yang, Jin Shao-Qin, Duan Ben-Song, Wu Shan, Chu Ye, Xiao Ding-Hua, Hu Li-Juan, Cao Jun-Zhi, Dai Jin-Feng, Liu Xiao, Xia Tian, Zhou Wei, Chen Tao, Zhou Chun-Hua, Wu Wei, Liu Shao-Jun, Yang Zhen-Yu, Wang Fen, Zhang Lu, Li Cheng-Zhong, Xu Hao, Wang Jun-Xue, Wei Bo, Lin Yong, Deng Xing, Qu Li-Hong, Shen Ying-Qiu, Wang Hui, Huang Yi-Fei, Bao Hai-Biao, Zhang Shuo, Li Li, Shi Yi-Hai, Wang Xiao-Yan, Zou Duo-Wu, Wan Xin-Jian, Xu Mei-Dong, Mao Hua, He Chao-Hui, Li Zhen, Zuo Xiu-Li, He Shui-Xiang, Xie Xiao-Ping, Liu Jun, Yang Chang-Qing, Spada Cristiano, Li Zhao-Shen, Liao Zhuan

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the diagnostic accuracy and safety of using magnetically guided capsule endoscopy with a detachable string (ds-MCE) for detecting and grading oesophagogastric varices in adults with cirrhosis.

DESIGN

Prospective multicentre diagnostic accuracy study.

SETTING

14 medical centres in China.

PARTICIPANTS

607 adults (>18 years) with cirrhosis recruited between 7 January 2021 and 25 August 2022. Participants underwent ds-MCE (index test), followed by oesophagogastroduodenoscopy (OGD, reference test) within 48 hours. The participants were divided into development and validation cohorts in a ratio of 2:1.

MAIN OUTCOME MEASURES: The primary outcomes were the sensitivity and specificity of ds-MCE in detecting oesophagogastric varices compared with OGD. Secondary outcomes included the sensitivity and specificity of ds-MCE for detecting high risk oesophageal varices and the diagnostic accuracy of ds-MCE for detecting high risk oesophagogastric varices, oesophageal varices, and gastric varices.

RESULTS

ds-MCE and OGD examinations were completed in 582 (95.9%) of the 607 participants. Using OGD as the reference standard, ds-MCE had a sensitivity of 97.5% (95% confidence interval 95.5% to 98.7%) and specificity of 97.8% (94.4% to 99.1%) for detecting oesophagogastric varices (both P<0.001 compared with a prespecified 85% threshold). When using the optimal 18% threshold for luminal circumference of the oesophagus derived from the development cohort (n=393), the sensitivity and specificity of ds-MCE for detecting high risk oesophageal varices in the validation cohort (n=189) were 95.8% (89.7% to 98.4%) and 94.7% (88.2% to 97.7%), respectively. The diagnostic accuracy of ds-MCE for detecting high risk oesophagogastric varices, oesophageal varices, and gastric varices was 96.3% (92.6% to 98.2%), 96.9% (95.2% to 98.0%), and 96.7% (95.0% to 97.9%), respectively. Two serious adverse events occurred with OGD but none with ds-MCE.

CONCLUSION

The findings of this study suggest that ds-MCE is a highly accurate and safe diagnostic tool for detecting and grading oesophagogastric varices and is a promising alternative to OGD for screening and surveillance of oesophagogastric varices in patients with cirrhosis.

TRIAL REGISTRATION: ClinicalTrials.gov NCT03748563.

Для цитирования: Jiang Xi, Pan Jun, Xu Qing, Song Yu-Hu, Sun Huan-Huan, Peng Cheng, Qi Xiao-Long, Qian Yang-Yang, Zou Wen-Bin, Yang Yang, Jin Shao-Qin, Duan Ben-Song, Wu Shan, Chu Ye, Xiao Ding-Hua, Hu Li-Juan, Cao Jun-Zhi, Dai Jin-Feng, Liu Xiao, Xia Tian, Zhou Wei, Chen Tao, Zhou Chun-Hua, Wu Wei, Liu Shao-Jun, Yang Zhen-Yu, Wang Fen, Zhang Lu, Li Cheng-Zhong, Xu Hao, Wang Jun-Xue, Wei Bo, Lin Yong, Deng Xing, Qu Li-Hong, Shen Ying-Qiu, Wang Hui, Huang Yi-Fei, Bao Hai-Biao, Zhang Shuo, Li Li, Shi Yi-Hai, Wang Xiao-Yan, Zou Duo-Wu, Wan Xin-Jian, Xu Mei-Dong, Mao Hua, He Chao-Hui, Li Zhen, Zuo Xiu-Li, He Shui-Xiang, Xie Xiao-Ping, Liu Jun, Yang Chang-Qing, Spada Cristiano, Li Zhao-Shen, Liao Zhuan. Diagnostic accuracy of magnetically guided capsule endoscopy with a detachable string for detecting oesophagogastric varices in adults with cirrhosis: prospective multicentre study.. BMJ (Clinical research ed.). 2024. DOI: 10.1136/bmj-2023-078581.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.